EasyManua.ls Logo

ORTEC 109A - Installation Instructions; Connection to Detector; Connection to a Shaping Main Amplifier; Input Power

ORTEC 109A
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.
INSTALLATION
INSTRUCTIONS
3.1
Connection
to
Detector
A
direct
connection
with
shielded
coaxial
cable
should
be
made
between
the
detector
and
the
BNC
connector
labeled
DETECTOR
on
the
front
panel.
The
performance
of
the
109A
Preampl
ifier,
like
that
of
all
other
such
low-noise
nuclear
amplifier,
is
degraded
as
the
capacity
at
the
input
of
the
amplifier
increases;
for
this
reason,
it
is
important
that
the
length
of
coaxial
cable
used
between
the
amplifier
and
the
detector
be
kept
at
the
minimum
necessary.
Also,
it
is
preferable
to
use
93-
or
100J2
impedance
cable
rather
than
75-
or
5012
cable,
since
the
capacity
per
foot
is
less
for
the
higher
impedance
cable.
Type
RG-62/U
cable,
which
has
9312
impedance
and
a
13.5-pF/ft
capacity,
is
recommended.
Type
UG-
260/U
connectors
fit
both
this
cable
and
the
BNC
input
connector.
Microdot
cables
and
fittings
of
the
series
stocked
and
suppl
ied
by
ORTEC
are
also
suitable.
The
cable
is
of
10012
impedance,
13
pF/ft.
An
adapter,
ORTEC
No.
C-17,
may
be
used
on
the
input
BNC
connector
to
permit
use
of
the
Microdot
cables.
See
ORTEC
1970
Nuclear
Price
List
for
a
complete
l
isting
of
compatible
Microdot
cables,
connectors,
and
adapters.
Once
the
input
cable
has
been
installed,
the
electronic
noise
performance
of
the
preamplifier
can
be
predicted
by
calculating
the
cable
capacity
from
the
above
information,
adding
the
capacity
expected
from
the
detector,
and
referring
to
the
table
of
typical
performance
versus
input
capacity
(Section
2).
3.2
Connection
to
a
Shaping
Main
Amplifier
The
preampl
ifier
can
be
used
to
drive
long
9312
l
ine
to
a
shaping
main
ampl
ifier
and
is
designed
to
be
directly
com
patible
with
the
ORTEC
transistor
main
ampl
ifiers.
It
can
be
used
with
any
shaping
main
amplifier
if
a
power
supply
is
used
to
power
the
preamplifier.
3.3
Input
Power
Power
for
the
109A
is
supplied
through
the
Amphenol
connector
(17-20090)
on
the
rear
of
the
chassis.
Power
may
be
supplied
by
a
single
45-V
battery
with
a
tap
at
22.5
V
(the
tap
is
used
as
ground,
providing
-t22.5
V
and
-22.5
V;
current
drain
is
36
mA),
or
any
well-filtered
power
supply
such
as
the
ORTEC
115
Preampl
ifier
Power
Supply
that
furnishes
both
+24
V
and
-24
V.
If
the
preampl
ifier
is
used
with
ORTEC
transistor
main
ampl
ifiers,
its
power
will
automatically
be
supplied
from
the
main
ampl
ifier
via
the
interconnecting
cable
suppl
ied
with
the
109A.
3.4
Test
Pulse
A
voltage
test
pulse
can
be
inserted
at
the
TEST
PULSE
connector
on
the
rear
of
the
109A
without
the
use
of
an
external
charge
terminator,
since
the
preamplifier
has
a
built-in
charge
terminator.
The
shape
of
this
voltage
pulse
must
have
a
fast
rise
(less
than
10"®
sec)
followed
by
a
slow
exponential
decay
back
to
the
basel
ine
(2
to4
x
10"^
sec).
The
input
ampl
itude
can
be
set
to
any
desired
level
with
the
knowledge
that
46
mV
amplitude
at
the
TEST
PULSE
connector
is
equal
to
approximately
1
MeV
energy
loss
in
a
silicon
detector.
Also,
the
test
pulse
can
be
inserted
into
the
DET.
INPUT
connector
simultaneously
with
an
operating
detector
by
using
an
external
charge
terminator,
provided
the
charge
terminator
will
withstand
the
detector
bias
voltage.
NOTE;
In
most
experimental
situations
the
optimum
signal-to-noise
ratio
occurs
with
the
preampl
ifier
gain
switch
in
the
XI0
position.
When
this
switch
is
in
the
XI
position,
the
signal-to-noise
ratio
at
the
preamplifier
output
is
the
same
as
that
of
the
XI0
position,
but
the
signal
level
is
only
1/10.
At
this
low
signal
level,
main
amplifier
noise
contribution
becomes
more
significant.
The
signal-to-noise
ratio
at
the
output
of
the
main
amplifier
can
be
degraded
by
the
amount
of
the
main
ampl
ifier
noise
contribution,and
resolution
somtimes
suffers.
4.
OPERATING
INSTRUCTIONS
4.1
Detector
Bias
Detector
bias
is
appl
ied
to
the
Preampl
ifier
through
the
high-voltage
SHV
connector.
NOTE:
The
detector
bias
components
and
connectors
are
rated
and
tested
at
1000
V.
Inside
the
109A
the
detector
bias
is
appl
ied
to
the
DET.
INPUT
connector
through
a
decoupling
network
consisting
of
R1
(1M12),
R2
(100M12)
and
C1
(1500
pF)
so
that
the
total
resistance
in
series
with
the
detector
bias
is
approximately
100M12.
This
resistance
must
be
taken
into
account
when
considering
the
bias
voltage
actually
appl
ied
to
a
leaky
detector.
The
voltage
dropped
in
this
network
wil
l
be
100
V
for
each
microampere
of
detector
leakage
current.
It
will
be
necessary
to
increase
the
bias
supply
output
voltage
to
compensate
for
this
drop.
Example:
It
is
desired
to
operate
a
detector
at
100
V
bias.
Detector
leakage
at
100
V
=
0.5
fxA.
The
drop
in
the
bias
network
=
(100
x
10®)
(0.5
x
10'®)
=
50
V.
Voltage
required
at
bias
supply
=
100
-I-
50
=
150
V.

Related product manuals