EasyManua.ls Logo

ORTEC 473 - Operation; Theory of Operation; General

Default Icon
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.
OPERATION
After
the
473
has
been
installed
and
interconnected
as
described
in
Section
3,
the
only
operating
functions
that
are
required
are
the
setting
of
the
detector
selector
switch,
adjustment
of
the
Disc
Level
control,
and
selection
of
the
desired
operating
mode.
f
Mi
ially
the
detector
selector
switch
can
be
set
at
the
switch
position
that
identifies
the
type
of
detector
being
used,
hovvever,
the
only
differences
between
the
three
switch
positions
marked
for
detector
types
are
the
effective
delay
time
that
will
be
used
for
constant
fraction
pickoff
and
the
controlled
dead
time
duration
in
the
output
circuit.
If
the
switch
is
set
at
Nal,
the
dead
time
for
each
output
pulse
will
be
1
fis
regardless
of
the
mode
by
which
the
output
was
generated.
The
controlled
dead
time
is
50
ns
for
any
of
the
other
three
switch
settings.
For
reference,
the
selectable
delay
times
are
14
ns
for
Ge(Li),
1.3
ns
for
Scint,
and
1.9
ns
for
Nal.
Shaping
that
is
provided
by
the
preamplifier
may
affect
the
relation
between
the
optimum
switch
selection
and
the
actual
class
of
detector
that
is
being
used.
If
the
switch
is
set
at
Ext,
the
delay
time
can
be
any
amount
greater
than
the
3
ns
built-in
delay
that
is
a
part
of
the
circuit
leading
to
and
from
the
rear
panel
connectors.
One
additional
function
is
selectable
when
the
Ext
setting
is
used,
and
this
is
the
effective
fraction;
the
basic
setting
of
a
jumper
on
the
printed
circuit
selects
30%,
but
the
jumper
can
be
moved
to
select
20%
or
10%
if
desired.
See
information
in
Section
5,
Theory
of
Operation,
for
changing
the
jumper
selection
of
the
effective
fraction.
The
function
of
the
Disc
Level
adjustment
is
to
permit
selection
of
signals
with
amplitudes
of
interest
and
to
eliminate
response
to
signals
with
smaller
amplitudes.
The
proper
setting
of
this
control
depends
on
the
range
of
signal
amplitudes
that
are
furnished
into
the
input.
The
control
range
is
from
50
mV
to
5
V,
with
the
10-turn
potentiometer
used
for
precise
setting
and
excellent
repeatability.
For
most
applications,
the
mode
selector
will
be
set
at
C.F.,
for
Constant
Fraction
operation.
If
the
input
signal
rise
times
tend
to
vary
through
a
wide
range,
such
as
the
signals
that
can
be
furnished
from
a
large-volume
coaxial
germanium
detector,
the
switch
can
be
turned
to
Slo
R.T.
Rej
to
operate
as
a
Constant
Fraction
discriminator
but
to
also
reject
input
pulses
with
too
slow
a
rise
time
so
that
the
timing
spectrum
has
a
better
resolution.
The
mode
switch
can
be
turned
to
L.E.
to
select
conventional
leading
edge
discriminator
operation.
If
the
detector
selector
switch
is
set
at
Nal,
the
internal
dead
time
is
held
at
1
(is
so
the
maximum
input
count
rate
is
limited
to
1
MHz.
But
if
the
detector
selector
switch
is
set
in
any
other
position,
the
dead
time
is
50
ns,
permitting
response
to
an
input
count
rate
up
to
20
MHz.
The
other
functions
selected
by
the
detector
switch
have
no
effect
during
L.E.
operation
because
the
constant
fraction
portion
of
the
473
circuits
is
bypassed.
5.
THEORY
OF
OPERATION
5.1.
GENERAL
The
circuits
of
die
473
are
shown
in
block
diagram
473-0101-81
and,
in
more
detail,
in
schematic
473-0101-S1.
Both
of
t
'.Gse
drawip'"^
are
included
at
the
back
of
the
manual.
The
input
is
furnished
simultaneously
to
three
internal
circuits.
One
circuit
is
a
lower
level
leading
edge
discriminator
that
defines
the
onset
of
a
signal
of
interest
and
arms
internal
logic
circuits
in
the
module.
A
second
circuit
is
an
upper
level
leading
edge
discriminator;
each
input
signal
must
exceed
this
level
in
order
that
an
output
will
be
generated.
The
third
circuit
is
a
constant
fraction
discriminator
that
provides
a
precise
timing
recognition
for
the
input
signal.
A
front
panel
Disc
level
control
effectively
adjusts
the
response
threshold
for
both
the
lower
level
and
upper
level
discriminators.
The
nominal
range
for
this
control
is
50
mV
to
5
V,
associated
with
the
upper
level
threshold
range;
the
threshold
for
the
lower
level
discriminator
is
automatically
adjusted
at
about
50%
of
the
upper
level
so
that
it
is
triggered
earlier
on
the
leading
edge
of
the
input
pulse.
When
an
output
is
generated,
it
furnishes
two
independent
NIM-standard
fast
negative
signals
and
one
NIM-standard
slow
positive
signal.
The
leading
edges
of
the
three
output
signals
are
coincident.
A
feedback
in
the
output
trigger
circuit
ensures
the
generation
of
only
one
output
pulse
in
each
of
the
three
output
signal
paths.
The
Constant
Fraction
mode
of
operation
can
be
used
either
with
or
without
the
Slow
Risetime
Reject
feature.
When
Slo
R.T.
Rej
is
selected
with
the
front
panel
mode
switch,
the
C.F.
timing
signal
is
not
generated
if
the
rise
time
of
an
input
pulse
is
too
slow.
The
Leading
Edge
mode
of
operation
does
not
use
the
slow
risetime
reject
feature.

Table of Contents