ORTEC 473 - Nai Mode Walk Test; Fig. 7.2. Test Setup for Scintillation Mode Walk Test

Default Icon
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17
HP
8004A
Rep
Rate
1K(CW)
Pulse
Width
lOnsiCCW)
Pulse
Del.
O(CCW)
Amplitude
5
V
Amp.
Vernier
-
CW
Pulse
Polarity
-
Neg.
DC
Offset
Off
Output
ThruSig.
In
\
Trigger
Output
TEK.
CT
3
TIME
PICK
OFF
1
TE
50
ATTENU
K
ATOR
Pick
Off
Stop
473
Shaping
Mode
-
Scint
Disc.
Mode
-
C.F.
Disc,
level
-
O.OSOV(CCW)
457
TIME
to
PULSE
HEIGHT
CONVERTER
Range
-
.05
CG
-
10
Multi
-
1
FG
-
(Adjust)
Bias
-
(Adjust)
Gate
-
Anti
Strobe
-
Int.
Pos.
Output
RG/62
-
(1
Meg)
1
p
100
'—wv—I
NOTE:
Alt
cables
are
SOU
coax
(RG/58)
unless
otherwise
specified.
TEK.485
SCOPE
Ext.
Trigger
Fig.
7.2.
Test
Setup
for
Scintillation
Mode
Walk
Test.
8.
Change
the
attenuator
in
the
following
sequence
while
observing
the
oscilloscope:
X5X10,
X10,
XI.
Adjust
R32
as
necessary
so
that
the
457
output
amplitude
changes
no
more
than
400
ps
(2
divisions)
for
the
attenuator
changes
in
step
8.
10.
Set
the
attenuator
at
X2X5X10
and
observe
the
peak
spread
of
the
457
output.
If
the
spread
exceeds
400
ps,
adjust
R32
to
reduce
it.
Repeat
steps
8
and
9.
If
R32
cannot
be
adjusted
to
satisfy
both
conditions,
replace
the
MC10116
integrated
circuit,
101.
7.5.
Nal
MODE
WALK
TEST
1.
Set
up
the
system
as
shown
in
Fig.
7.3.
Adjust
the
HP8004A
output
pulse
width
to
25
ns.
2.
Use
X2X5
on
the
attenuator.
3.
Adjust
the
457
bias
so
its
output
pulse
shows
about
6
V
of
amplitude
as
seen
on
the
oscilloscope.
4.
Set
the
vertical
gain
of
the
oscilloscope
for
1
V/division.
5.
Adjust
the
457
bias
and
Fine
Gain
until
a
change
of
1
ns
corresponds
to
a
IV
(1
-division)
change
on
the
oscilloscope.
6.
Change
the
oscilloscope
vertical
gain
to
0.5V/division.

Table of Contents