5 Technical specifications
Standards compliance: EN 60645-1: 1996
EN 60601-1:1990 + A1:1993 + A2: 1995 + A13:1996
EN 60601-1-2:2002
Classification: Group 1, class A EN 60601-1-2:2002
Medical CE- mark: Inmedico A/S is approved for medical CE marking, by DGM.
Identification number 0543
Transducers: TDH-39 air, B-71 bone (calibrated via mastoid)
Power Supply: 18 VAC, 800mA max.
Adaptor: Primary: 230 VAC, 50/60 Hz, 170mA max.
Secondary: 18 VAC, 20 VA, ± 10 %
Environmental Conditions for Operation
Ambient Temperature: +15 to +35 degree Celsius.
Relative Humidity: 30 % to 90 %
Surrounding pressure: 80 kPa to 120 kPa
Environmental Conditions for Storage
Ambient Temperature: -10 to +50 degree Celsius
Relative Humidity: 95% or less (non-condensing)
Surroundings pressure: 50 kPa to 120 kPa
Physical Attributes
Dimensions: 389 (W) x 180 (D) x 55 (H) mm
Parallel port: Automatic printout of complete audiogram, 60/72 dpi.
Serial port
: Computer data transmission (RS-232) to patient file. 9600 baud.
Warm-up time: < 10 minutes
Included parts: audiometer, Peltor H7A headphones with TDH-39, response button,
bone conductor, power supply, user manual and calibration certificate
Accessories: carrying bag, serial cable, printer cable, AudioConsole software
Measurement specifications
Measurement Method Manual and automatic threshold test with storage of data.
Type of protection against electric shock: Class I equipment
Degree of protection against electric shock: Type B applied part
Degree of protection against liquid penetration: IPO, ordinary equipment
Degree of safety of application in the presence
of flammable anaesthetics:
N/A
Mode of operation: Continuous operation