ID: 1614 / ver. 210
selects the next mode and DOWN selects the previous. The mode is displayed breefly when changed, and the pulse
and response counters are reset. The test can be terminated manually with the interrupter.
The memory:
Test results are stored in every audiogram, and for every frequency as a percent figure. There is memory for the left
and the right earphone. It is possible to run a SISI test using the bone conductor and the mono insert phone too, but
the results are not stored separately, they use the same memory as left and right. When SISI is selected on the channel
2 function selector, the VU display recalls the stored test results when the frequency and channel is changed. These
results may be erased by pressing CLEAR (with shift). This only clears the single result that corresponds to the currently
selected frequency and channel. When a printout is done, all the SISI test results are shown in a table below the
audiogram.
ABLB: (Alternate Bilateral Loudness Balance)
ABLB involves both channels by activating the two interrupters twice on each channel, in the following way:
INPUT and OUTPUT is selected for both channels. The two attenuators are set on the wanted soundpressure. On the
channel 2 FUNCTION choose ABLB.
To start the test: press the channel 2 interrupter.
Duration of tone as well as the pause in between may be programmed individually, SETUP no. 7.
4.16 Setup
You may individually change the basic function of the SM960-C by programming one or more of the 11 SETUP
functions.
The setup function is entered by
pressing the SETUP button.
Two short beeps will be heard, and the
VU display will show “SETUP”
briefly.
To change between each SETUP mode press the channel 2 attenuator, and selected mode is shown in the channel 2
attenuators display as a figure between 1 and 11.
Mode 1: Frequency selecting
AUDIOGRAM ID
FRE 125-8000 1
FREQUENCY VU/VOLUME ATTENUATOR 2
In this mode you may leave out one or more frequencies. Select the frequency with the frequency buttons, and use
the channel 1 attenuator to include or leave out the selected frequency. The VU display says ON when the frequency
is included and OFF if it is left out.
Mode 2: Tone length
AUDIOGRAM ID
LENG OFF 2
FREQUENCY VU/VOLUME ATTENUATOR 2
The duration of the ON-TIME of the interrupter. To change the ON-TIME, use the channel 1 attenuator. The selected
length is shown in the VU display. Options are: OFF, 0.5, 1.0, or 1.5 second.
If one of the three numbers are programmed the tone will last for this period, no matter how long you press the
interrupter. If the display says OFF, the interrupter is operated manually.
Mode 3: Recall mode