ID: 1614 / ver. 210
AUDIOGRAM ID
RCAL ON 3
Connects or disconnects the storage function. To include or exclude storage capacity press the channel 1 attenuator.
The VU display will show ON or OFF for connected or disconnected. If the memory function is disconnected the device
will operate as a standard manual audiometer, and printer, serial port and automatic tests are not accessible. A special
setting here is “AUTO”. If this is selected, the audiometer will transmit the audiogram to the serial port each time the
channel 1 attenuator is changed. It also transmits when the ID number is changed, and when an automatic test is run,
each time the frequency is changed. Note that the channel 1 attenuator will be acting a bit slow while this “AUTO”
transmit feature is enabled.
Mode 4: Reset
AUDIOGRAM ID
RSET OFF 4
FREQUENCY VU/VOLUME ATTENUATOR 2
Connects or disconnects the RESET option. The programming of this feature is done by means of the channel 1
attenuator, and the VU display says ON or OFF. When reset is ON the tone is disconnected when you change
frequency or channel.
Mode 5: Auto mode
AUDIOGRAM ID
PRNT HP 5
FREQUENCY VU/VOLUME ATTENUATOR 2
By means of programming mode 5 you may choose the type of printer which is connected to the device. Selection is
made by pressing the channel 1 attenuator. The selected printer type is shown in VU display. Options are:
9: Used with most Epson compatible 9 pin matrix printers.
24: Used with most 24 pin printers, and many inkjet printers such as Canon.
PRO: Used with IBM ProPrinters. Some other printers can emulate a ProPrinter.
HP: Used with HP DeskJet and laser printers. Many other laser printers may emulate HP lasers, and will work with this
setting.
Note that printers designed specially for Microsoft Windows will not work when connected directly to the audiometer.
It is, however, perfectly possible to use them together with a PC, when the audiometer is connected to the PC via the
serial port. If you don’t know which ‘language’ your printer works with, you may just try the different settings and see
if it works. There is no risk of damage to neither the printer nor the audiometer by selecting a wrong printer type. You
may also consult the printers’ manual for information about what the printer can emulate.
Mode 6: Auto mode
AUDIOGRAM ID
AUTO RND 6
FREQUENCY VU/VOLUME ATTENUATOR 2
Selecting one of the three automatic tests. In the VU display is shown which of the three automatic tests is selected:
20dB, Random 20dB (RND) or variable (VARI) and the choice is made by means of the attenuator 1 buttons. Please
note that any frequencies left out as a result of programming mode 1, will also be skipped over in the AUTO test.
Mode 7: ABLB test timing
AUDIOGRAM ID
ON 0.6 7