14
FREQUENCY VU/VOLUME MASKING LEVEL dB
Programming of ON and PAUSE time for the ABLB test. Select between ON and PAUSE time with the channel 1
interrupter button. The FREQUENCY display will show ON = ON TIME or PAUS = PAUSE. Then use the channel 1
attenuator to select a time between 0.1 and 1 second.
Mode 8: Carriage return on/off
AUDIOGRAM ID
CR ON 8
FREQUENCY
VU/VOLUME
MASKING LEVEL dB
Some printers may not print a full audiogram, or make blank horizontal stripes. This is usually caused by either
sending a carriage return to a printer programmed to automatically return the printhead at a new line (blank stripes
occur), or by sending no carriage return to a printer that expects it (only one line is printed). By means of the channel
1 attenuators you may connect or disconnect carriage return. The VU display will say ON or OFF for connected or
disconnected. This setting has no effect if the printertype (mode 5) is set to HP.
Mode 9: Bone audiogram on/off
AUDIOGRAM ID
BONE ON 9
FREQUENCY
VU/VOLUME
MASKING LEVEL dB
On the printout you may choose to have the bone test results printed in a separate audiogram frame, or have both air
and bone printed in the same frame. By selecting ON in this mode, a separate bone audiogram is printed. If this mode
is on, and no air tests have been performed, only a single audiogram is printed, containing the bone test results. Also,
if no bone results exist in memory, no extra audiogram is printed.
Mode 10: Identification on/off
AUDIOGRAM ID
ID
ON
10
FREQUENCY VU/VOLUME MASKING LEVEL dB
This mode is for enabling or disabling the identification sent to the PC when turning on the device. This must be
turned on for use with the AudioConsole program. If you encounter problems when the device is used to transmit
data to other programs, try turning ID off.
Mode 11: Select supra phones or inserts phones calibration
AUDIOGRAM ID
CALI PHON 11
FREQUENCY VU/VOLUME MASKING LEVEL dB
This mode selects between supra aural earphones or insert phones. The selected type of phones is displayed briefly
when the device is turned on. Select “PHON” if supra aural phones (TDH 39) are used, and “INST” if insert phones
(usually Eartone 3A or 5A) are used. The selection affects the calibration and the maximum output level.
Mode 12:
AUDIOGRAM ID
PP PARA 12
FREQUENCY VU/VOLUME MASKING LEVEL dB
This mode selects if a printer is connected to the parallel port, or the serial port. If you use the Kyosha Thermo printer,
this mode must be set to SERI. This printer is connected to the socket marked “SERIAL” on the audiometer. For
printers with parallel port, select PARA, and connect the printer to the socket marked “PRINTER”.
Finishing programming