17
Medium Priority Alerts (Continued)
!
C
Description Display Icons Condition/Action/Explanation
Battery HOT
Warning
Battery has exceeded temperature limit while
concentrator is running on battery power.
If possible, move concentrator to a cooler
location or power unit with an external power
supply and remove battery. If condition
persists, contact your equipment provider.
System Hot
Warning
Concentrator temperature has exceeded
temperature limit. If possible move
concentrator to a cooler location. Ensure air
intake and outlet vents have clear access and
particle lters are clean. If condition persists,
contact your equipment provider.
High Priority Alerts
The following high priority alerts are accompanied by a ve beep pattern, repeated every
10 seconds and a ashing yellow light.
Description Display Icons Condition/Action/Explanation
Battery Empty
Attach Plug
Concentrator has insucient battery power to
produce oxygen. Attach external power supply
or exchange battery, then restart unit if
necessary by pressing On/O button.
Battery HOT Battery has exceeded temperature limit while
concentrator is running on battery power.
Concentrator has stopped producing oxygen.
If possible, move concentrator to a cooler
location, then turn power o and back on.
Ensure air intake and outlet vents have clear
access and particle lters are clean. If condition
persists, switch to a backup source of oxygen and
contact your equipment provider.
System HOT Concentrator temperature is too high and
oxygen production is shutting down. Ensure air
intake and outlet vents have clear access and
particle lters are clean. If condition
persists, switch to a backup source of oxygen
and contact your equipment provider.