EasyManua.ls Logo

Packo HP64 - Page 15

Default Icon
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
Pumped fluid changes state in contact
with air, i.e. crystallizes, forms a film,
dries out or precipitates solids. The flush
dissolves and rinses away the small
amount of residue which could build up on
the edges of the seal faces.
Pumped fluid is close to its boiling point,
e.g. water over 80°C (175°F). The
flushing fluid is used to cool the seal faces.
Pumped fluid is temperature sensitive and
evaporates when heated by the shearing
action of the seal faces. The flushing fluid
is used to cool the seal faces.
Pump may run “dry”, i.e. no liquid in
pump chamber for very short periods.
Pump is under high vacuum.
A low pressure sterile barrier is required.
A low-pressure flushing fluid system must
be installed as follows:
Liquid must be compatible with the
pumped fluid; water is the most
commonly used liquid.
Pressure shall typically be 0.5 bar (7 psi)
gauge. Maximum of 1.0bar (14.5 psi)
gauge.
Temperature shall be maximum of 70ºC
(160ºF) for water, less for volatile liquids
(minimum 20ºC (68ºF) below boiling point
of liquid).
Flow rate shall preferably be 2 to 3
litres/min. (0.5 to 0.75 US gal/min) per
seal.
Flush fluid should be connected to enter
at the lowest point on the seal housing
and flow out at the highest point to vent
air pockets.
Pipework should be as shown in the
diagram (Figure 11).
Figure 11 - Suggested low pressure flushing system
The Double Mechanical Seals fitted to HP &
LH pumps provide the facility to contain a high
pressure fluid behind the primary seal. This allows
the pump to be used for applications where the
single seal is unsuitable.
Double seals code D are run with a fluid between
the primary and secondary mechanical seals to
form a barrier between the pump and the
atmosphere. (Figure 12)
Figure 12 - Double mechanical seal (Size 4-8)
Double Mechanical Seal with a low pressure
liquid flush when:
ATTENTION
The pumped fluid is toxic or hazardous
and must not escape from the pump even
in minute quantities.
The system can be installed as in figure 12,
but must be capable of withstanding the

Related product manuals