Medical Infrared Thermometer
PC809 / PC809EV / PC809V
Version 2.0
2019.09.01
Users Manual

Questions and Answers

  • A
    Andrea FarrellAug 5, 2025
    How to fix PACOM PC809 that failed to power on?
    • J
      Jerry ChavezAug 5, 2025
      If your PACOM Thermometer fails to power on, there are a few potential causes. First, the battery level might be too low, in which case you should replace the batteries. Second, the polarities of the batteries might be reversed; ensure that the batteries are inserted correctly. If neither of these resolves the issue, the thermometer may be damaged, and you should contact the dealer or manufacturer.
  • A
    Alan ScottAug 12, 2025
    What to do if PACOM PC809 reading is too high?
    • J
      Jessica PhillipsAug 12, 2025
      If your PACOM Thermometer displays a reading that is too high, consider these possibilities: you may have just come from a hot environment, or you are exposed to hot air. In such cases, stay in a room between 10.0°40.0 ° (50.0°104.0°) for at least 30 minutes before taking a temperature measurement. Also, a low battery can cause this issue, so try replacing the batteries.
  • T
    teresa52Aug 17, 2025
    Why PACOM PC809 Thermometer reading is too low?
    • P
      Peter RomeroAug 17, 2025
      If your PACOM Thermometer shows a reading that is too low, it could be due to several reasons. The infrared sensor lens of the probe might be dirty; clean it with cotton swabs dipped in anhydrous ethanol. Alternatively, the distance between the probe and the target might be too great; ensure the probe touches your forehead or is placed in your ear canal. Finally, if you have just come from a cold environment, stay in a room between 10.0°40.0 ° (50.0°104.0°) for at least 30 minutes before taking a temperature measurement.
  • J
    Jeremy MejiaAug 23, 2025
    What to do if PACOM PC809 Thermometer shows 'Hi'?
    • R
      Renee JordanAug 23, 2025
      If your PACOM Thermometer displays 'Hi', it indicates that the human body temperature is over 42.9°/109.2°. First, ensure the probe is clean. If the environmental temperature is out of range, both the patient and thermometer should stay in a room between 10.0°40.0° (50.0°104.0°) for at least 30 minutes before taking a temperature measurement. If the problem persists, contact the dealer or manufacturer.
  • L
    Linda DunnAug 29, 2025
    What does 'Lo' mean on PACOM PC809 Thermometer?
    • A
      Amber NguyenAug 30, 2025
      If the PACOM Thermometer displays 'Lo', it means the human body temperature is below 32.0°/89.6°. First, make sure that the probe is clean and free from dirt. Second, if the environmental temperature is out of range, then both patient and thermometer should stay in a room between 10.0°40.0° (50.0°104.0°) for at least 30 minutes before taking a temperature measurement. If the problem continues after these steps, contact the dealer or manufacturer.

Summary

Introduction

Packing List

Product Description

Product Overview

Describes the thermometer's function and normal temperature ranges for different body sites.

Intended Use and Principle

Explains the device's purpose and how it measures temperature using infrared energy.

Features, Components, and Structure

Details the thermometer's capabilities, parts, and physical structure with diagrams.

LED Display Explanation

Explains the meaning of symbols and indicators on the LED display.

Warnings and Precautions

Directions for Use

Forehead Temperature Measurement

Instructions for taking temperature from the forehead, including probe placement and button press.

Ear Temperature Measurement

Steps for taking temperature from the ear, including probe insertion and ear canal preparation.

Recall Saved Readings

How to access and review previously stored measurement data.

Temperature Unit Conversion

Instructions for switching between Celsius and Fahrenheit temperature units.

Audio;Vibration Feedback Settings

How to enable or disable the beep, speaker, or vibration alert.

Low Fever Threshold Adjustment

Procedure to manually adjust the low fever alarm temperature setting.

High Fever Threshold Adjustment

Procedure to manually adjust the high fever alarm temperature setting.

Device Shutdown

Instructions on how to turn off the thermometer, automatically or manually.

Battery Replacement

Guide on how to replace the batteries when the low battery indicator appears.

Temperature-Taking Tips

Cleaning and Disinfection

Cleaning Procedure

Instructions for cleaning the thermometer shell and the infrared sensor lens.

Disinfection Procedure

Guidelines for disinfecting the thermometer body and probe using medical alcohol.

Troubleshooting

Specifications

Symbol Descriptions

Maintenance

After-sales Service

Declaration

Appendix A: EMC Information

Manufacturer Information

PACOM PC809 Specifications

General IconGeneral
Product nameMedical Infrared Thermometer
Product modelPC809/PC809EV/PC809V
Applicable regulations and lawsASTM E 1965 / EN12470-5 / GB/T 19146
Date of manufactureSee label
Life5 years
Power supplyDC 3.0V (2 × AAA batteries)
Automatic shutdown30s±1s
Battery lifeMore than 2, 000 measurements
Low battery indicatorAppears on the screen when the voltage is below 2.6V±0.1V
Measurement modesForehead Mode and Ear Mode
Measurement time1 second
Measurement range32.0‒42.9℃ (89.6‒109.2℉)
Accuracy±0.2℃/0.4℉
Temperature units℃ and ℉ (convertible)
Operation environmentTemperature: 10.0‒40.0℃ (50.0‒104.0℉) Humidity: 20‒85% RH, non-condensing Atmospheric pressure: 86‒106kPa
Storage and shipping environmentTemperature: -20‒55℃ (-4‒131℉) Humidity: 20‒93% RH, non-condensing Atmospheric pressure: 86‒106 kPa
Display resolution0.1℃/℉
Memory32 saved reading slots
Weight and Dimensions IconWeight and Dimensions
Net weight70 g (not including batteries)
Dimension155 × 38 × 45 mm