Keep the infrared sensor lens away from water during the
cleaning process, or else it may be damaged.
The infrared sensor lens may be scratched if it is cleaned with a
piece of tissue paper, resulting in inaccurate readings.
Do not allow the infrared sensor lens to make contact with hard
objects.
Never use abrasive cleaning agents, thinners or benzene for
cleaning.
Never immerse any part of the thermometer in liquid, or allow
liquid to enter the device.
7.2 Disinfection
Disinfect the thermometer body and the area around the probe
with a soft cloth slightly moistened with 75% medical alcohol. Do
not use the same method to disinfect the infrared sensor, as 75%
medical alcohol contains water which may damage the lens.
Do not use hot steam or ultraviolet radiation for disinfection, or
else the thermometer may be damaged or quickly aged.
!
!
!
!
!
!
- 15 -