Display | LCD |
---|---|
IP Rating | IP67 |
Light Source | LED |
Operating Temperature | 0 - 50 °C |
Storage Temperature | -10 - 60 °C |
Waterproof Rating | IP67 |
Parameters Measured | Free Chlorine, Total Chlorine, pH, Alkalinity, Cyanuric Acid |
Method | Colorimetric |
Range | Free Chlorine: 0 - 5 mg/l, Total Chlorine: 0 - 5 mg/l, pH: 6.5 - 8.4 pH, Alkalinity: 0 - 500 mg/l, Cyanuric Acid: 0-100 mg/l |
Resolution | 0.01 mg/L (Chlorine), 0.1 pH, 1 mg/L (Cyanuric Acid), 1 mg/L (Alkalinity) |
Battery Life | Approximately 1000 tests |
Essential steps and tips for obtaining accurate test results with the Lumiso photometer.
Using Palintest Photometer tablets, and avoiding rapid dissolving or Comparator tablets.
Rinsing the test tube with sample, leaving a few drops, for chlorine testing.
Adding a DPD1 or DPD XF tablet to the sample in the test tube.
Crushing the DPD tablet to form a paste within the test tube.
Adding one Oxystop tablet to the sample for Total Chlorine measurement.
Crushing and stirring the contents after adding the Oxystop tablet.
Replacing the cap and waiting 1 minute before proceeding with Total Chlorine measurement.
Filling the test tube to the 10mL line with the sample for pH testing.
Adding one Phenol Red tablet to the sample in the test tube.
Crushing the Phenol Red tablet and stirring the mixture to dissolve it.
Filling the test tube to the 10 mL line with sample for Cyanuric Acid testing.
Adding one Cyanuric Acid tablet without crushing for the test.
Waiting 2 minutes for the Cyanuric Acid tablet to disintegrate before proceeding.
Filling the test tube to the 10 mL line with sample for Total Alkalinity testing.