Kemio
User Manual

Questions and Answers

  • S
    sarah80Sep 8, 2025
    What to do if Palintest Test Equipment displays 'No barcode found'?
    • F
      Frank GloverSep 9, 2025
      If your Palintest Test Equipment says 'No barcode found', it means the scanner couldn't detect the barcode. Try scanning the barcode again within 5 seconds.
  • J
    James EllisAug 23, 2025
    What to do if palintest Kemio shows 'Outside temperature spec'?
    • W
      William FigueroaAug 23, 2025
      If your palintest Measuring Instruments displays 'Outside temperature spec', it means the sample temperature is outside the sensor's specified performance range. While a result is displayed, it may be less accurate. Bring the sample to within the specified temperature before testing to ensure accurate results.
  • H
    Heather GonzalezAug 25, 2025
    What to do if palintest Kemio displays 'Invalid Barcode'?
    • R
      Richard Rodriguez PhDAug 26, 2025
      If your palintest Measuring Instruments displays 'Invalid Barcode', it means the scanned barcode does not match the Kemio barcode format. Make sure to scan the correct barcode type to resolve this issue.
  • A
    Alexander Mahoney DDSAug 27, 2025
    What to do if palintest Kemio displays 'Invalid Sensor Type'?
    • M
      Melinda WadeAug 27, 2025
      If your palintest Measuring Instruments displays 'Invalid Sensor Type', it indicates that the scanned Kemio sensor does not match the selected test and its limits. To fix this, scan a barcode from a suitable sensor batch.
  • R
    robinmendozaAug 27, 2025
    What to do if palintest Kemio displays 'User aborted'?
    • R
      raymondwoodAug 28, 2025
      If your palintest Measuring Instruments displays 'User aborted', it means the user pressed cancel or home during the test. Repeat the test with a new sensor and do not interrupt the test to avoid this issue.
  • R
    Richard BoydSep 22, 2025
    What to do if Palintest Test Equipment says 'No test selected'?
    • K
      Karen HerreraSep 22, 2025
      If your Palintest Test Equipment indicates 'No test selected', it means all test types are unselected. To resolve this, select at least one test type.
  • C
    Chelsea CooperSep 16, 2025
    What to do if Palintest Test Equipment is 'Outside temperature spec'?
    • D
      Dominique MaddenSep 16, 2025
      If your Palintest Test Equipment warns about being 'Outside temperature spec', it means the sample temperature is outside the specified performance range of the sensor. While a result is displayed, it may be less accurate. Bring the sample to within the specified temperature before testing.
  • G
    Gabriel MartinezSep 18, 2025
    How to fix 'Start up electrode leakage' on Palintest Test Equipment?
    • B
      bharrisSep 18, 2025
      If your Palintest Test Equipment shows 'Start up electrode leakage', it could be due to a couple of reasons: either the sensor contacts are wet, or a sensor was left in the instrument during start up. Here's what to do: 1. Open the front lid of the arm and wipe the contacts with a dry cloth. 2. Discard the sensor and sample after testing, before powering down the instrument.
  • G
    Gina BeckSep 7, 2025
    What causes 'Invalid Sensor Type' error on Palintest Test Equipment?
    • P
      pcohenSep 7, 2025
      If your Palintest Test Equipment displays 'Invalid Sensor Type', it means the scanned Kemio sensor doesn't match the selected test and its limits. To correct this, scan a barcode from a suitable sensor batch.
  • H
    Holly MorganSep 4, 2025
    What to do if Palintest Test Equipment shows 'Invalid Barcode'?
    • D
      diazwayneSep 4, 2025
      If your Palintest Test Equipment displays 'Invalid Barcode', it means the scanned barcode doesn't match the Kemio barcode format. To fix this, ensure you are scanning the correct barcode type.

Summary

About Palintest

Palintests History of Innovation

Palintest is committed to making water analysis technologies simple and accessible.

Palintests Global Presence

Palintest operates globally, with a presence in the USA, UK, China, and Australia.

Kemio Introduction

Kemio Product Overview

Provides a visual overview and lists key components of the Kemio instrument.

Kemio Safety Information

Advises users to read the manual before setup and operation for optimal performance.

Kemio Kit Contents

Benchtop Kit Contents

Lists the items included in the Kemio Benchtop Kit, suitable for static applications.

Soft Kit Contents

Lists the items included in the Kemio Soft Kit, designed for mobile applications.

Chlorite Hard Case Kit Contents

Lists the items included in the Kemio Chlorite Hard Case Kit for chlorine testing.

Performing a Kemio Test

Kemio Test History

Benefits of Kemio Configuration

Kemio Data Enrichment

Explains how tagging information adds context and value to test results.

Kemio Labeling for Tests

Describes the use of labels for tagging tests and providing context.

Kemio User Profiles

Explains how to create user profiles for accountability and traceability.

Kemio Reference Codes

Kemio Traceability Features

Improve Accountability with Users

How using Kemio to tag users improves accountability.

Kemio Batch Code Scanning

Ensures correct sensor batch information is used by scanning before each test.

Kemio Pass;Fail Results

Kemio Process Control

Kemio Actionable Results

How PASS/FAIL results enable rapid decision-making and corrective actions.

Kemio Dilution Correction

Automatic correction for dilution factors programmed into labels.

Simple Kemio Configuration

Selecting Kemio Test Types

How to select a single target test parameter for simpler operation.

Kemio Label;User Configuration Options

Options for disabling or simplifying the tagging of labels, limits, and users.

Kemio Barcode Scanning Settings

Configuring when Kemio prompts users to scan sensor barcodes.

Kemio Configuration Menu

Managing Kemio Labels

Explains how labels are used for data enrichment, traceability, and process control.

Adding a Kemio Label

Step-by-step guide to adding new descriptive labels with optional limits.

Kemio Label Sample Configuration

Editing Kemio Labels

Deleting Kemio Labels

Kemio User Management

Adding a Kemio User

Steps to add a new user profile for performing tests.

Deleting a Kemio User

How to remove user profiles from the instrument.

Kemio Batch Information

Adding a Kemio Batch

Guide to adding sensor batch information via the configuration menu.

Scanning Kemio Batch QR Codes

Viewing Kemio Batch Details

Kemio Check Standards

Validating Kemio with Check Standards

Step-by-step guide on how to perform check standards tests.

Selecting Kemio Test Types

How to Select Kemio Tests

Detailed steps for selecting desired test types within the configuration menu.

Kemio Settings Menu

Kemio Display and Input Settings

Covers date format, date, time, language, and units settings.

Kemio Test Configuration Settings

Includes test settings, user, labels, and reference configuration.

Kemio Scanner Settings

Configuration for the barcode scanner to ensure correct sensor type and calibration.

Kemio Scanner Settings Details

Viewing Kemio Instrument Details

How to access instrument details for troubleshooting.

Kemio Calibration and Validation

Kemio Batch Information Details

Importance of using correct batch information for sensor calibration.

Kemio Check Standards Function

Kemio Data Connectivity

Transferring Kemio Data

Steps to connect Kemio to a PC and navigate to view test logs and results.

Kemio Data Log Explained

Describes the structure and content of the Kemio data log.

Kemio Care and Maintenance

Cleaning the Kemio Sample Vessel

Instructions for rinsing and emptying the sample vessel to prevent contamination.

Kemio Sensor Handling

Advice on handling, storing, and disposing of single-use Kemio sensors.

Kemio Check Standards Care

How to keep check standards clean and dry, and their expiry date.

Kemio Degassing Equipment

Replacing Kemio Degasser Filter

Step-by-step guide to replacing the filter in the degassing equipment.

Kemio Battery Information

Kemio Warranty Information

Kemio Troubleshooting Guide

Kemio Barcode and Sensor Errors

Covers invalid barcode and invalid sensor type errors.

Kemio Batch and Test Selection Errors

Covers invalid batch and no test selected errors.

Kemio Instrument Handling Errors

Covers instrument rolled and lid opened during test errors.

Kemio Technical Support

Kemio Reorder Codes

Kemio Technical Specifications

palintest Kemio Specifications

General IconGeneral
Operating Temperature0 - 50 °C
IP RatingIP67
WaterproofYes
Measurement ParametersFree Chlorine, Total Chlorine, Chlorine Dioxide, Bromine, Ozone, Peracetic Acid, Hydrogen Peroxide
Resolution0.01 mg/L
DisplayLCD
ConnectivityBluetooth
Data Storage1000 results