6.7 Troubleshooting
The Potalab
®
+ incubator monitors performance
during all stages of performance and provides
information regarding unexpected condition
as follows:
Error 102: Low Temp
The incubator has failed to reach the temperature
set point within the specified time limit. Remove
from areas of low
temperature/high cooling and
accept the error.
Error 103: High Temp
The incubator temperature has exceeded the set-
point for a significant period. Remove from direct
sunlight or high temperature conditions and
accept the error.
Error 107: Validation
The internal validation of thermistor temperature
monitoring has exceeded the tolerance of the
device. Return to authorised
service centre for
attention as soon as possible.
Error 110: Battery Low
The 12V DC battery has insufficient battery capacity
to complete the full incubation cycle. Prepare
alternative power supply as soon as possible.
Error 111: Battery Critical
The 12V DC has approximately 10 minutes
battery life remaining. Find alternative power
source immediately.
Error 112: Power Loss
Unexpected power loss has interrupted incubation.
Check for cause if possible and restart incubation
when reliable power is available.
A full list of error codes is found
in Appendix 2 - Troubleshooting.
Incubator Contamination
Should any hazardous material be spilt onto or
into the incubator, cleaning and decontamination
should only be carried out with a damp cloth
and mild detergent.
Disconnect power during cleaning and do not
invert the incubator.
Do not immerse the incubator under
any circumstance.
Do not use acetone or any abrasive/aggressive/
hazardous cleaning agents. For advice
regarding acceptable cleaning agents contact
support@palintest.com.
User Service
The Potalab
®
+ incubator contains no user
serviceable parts. The automated calibration
validation will confirm the incubator is
operating correctly.
For service please contact your local
Palintest representative or email to
support@palintest.com.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the palintest Wagtech Potalab+ (M) and is the answer not in the manual?

palintest Wagtech Potalab+ (M) Specifications

General IconGeneral
Brandpalintest
ModelWagtech Potalab+ (M)
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals