Questions and Answers

  • J
    Justin JimenezAug 18, 2025
    What to do if the outer door of my Panasonic MCO-230AIC incubator does not open?
    • J
      Jessica CharlesAug 19, 2025
      When the power switch is OFF, the electric lock is activated, preventing the outer door from opening. To resolve this, either turn the power switch ON or use the accessory unlock key to override the electric lock. Also, the outer door is electrically locked during decontamination and cannot be opened.
  • A
    alexandra56Aug 21, 2025
    Why is CO gas not being injected into my Panasonic MCO-230AIC Accessories incubator?
    • C
      Christopher SchultzAug 21, 2025
      The CO control method for the incubator is the ON-OFF method. CO gas is intermittently injected as the gas density in the chamber approaches the set value. Also, the gas is not injected until the temperature of the CO sensor becomes stable enough, approximately 1 hour, after turning ON the power switch or recovering from power failure.
  • R
    Richard GreenAug 26, 2025
    Why is the CO gas density slow to recover in my Panasonic MCO-230AIC Accessories incubator?
    • R
      Robin WilsonAug 26, 2025
      The CO gas density may be slow to recover due to the following reasons: * A HEPA filter is used for the incubator CO gas piping. If gas density is slow to recover when the CO gas pressure is normal, it is possible that the HEPA filter may be clogged. * There is little gas remaining in the CO gas cylinder. * The secondary pressure for the gas regulator is not at the specified value of 0.03 MPa(G) (0.3 kgf/cm2(G), 4.4 psi(G)). * The gas tube is blocked. * The duct is not securely attached. Attach the duct properly to the 4 points hooks. * The fan is not attached properly. Confirm if the fan is pushed all the way to the motor shaft.
  • W
    willie88Aug 27, 2025
    What causes the chamber temperature to not match the set value in my Panasonic MCO-230AIC incubator?
    • O
      onealannaAug 28, 2025
      The chamber temperature may not match the set value if the ambient temperature is less than 5 °C different from the set value, the outer door is closed but the inner door is left open, or the incubator is operating beside an appliance that generates electromagnetic waves.
  • L
    Lauren HinesAug 30, 2025
    Why is a large quantity of CO gas being consumed by my Panasonic MCO-230AIC Accessories incubator?
    • J
      Joshua BentleyAug 30, 2025
      A large quantity of CO gas may be consumed if the outer and inner doors are being frequently opened and closed, gas is leaking from connectors due to deterioration of the gas tube or pinhole leaks, the packing seal for the inner door is defective, or the access hole is open.
  • K
    kevin46Sep 1, 2025
    Why H O decontamination cannot be performed in my Panasonic Accessories incubator?
    • J
      Jamie HernandezSep 1, 2025
      If the MCO-230AICUVH is being used, check if the optional MCO-HP is installed. If the MCO-230AICUV is being used, check if the optional MCO-HP, the optional MCO-170HB and the MCO-170EL are installed. If the MCO-230AIC is being used, check if the optional MCO-170UVS, the MCO-HP, the optional MCO-170HB and the MCO-170EL are installed. Ensure the UV lamp is not burned out. Verify the H O generator cable is properly connected and that the entire bottle of the H O reagent has not been used.
  • C
    Christopher DanielSep 4, 2025
    What to do if the chamber humidity does not rise in Panasonic MCO-230AIC?
    • A
      Ashley GarciaSep 4, 2025
      The chamber humidity may not rise if there is not enough water in the humidifying pan. Be sure to use sterile distilled water.
  • C
    Chelsea HarrisSep 6, 2025
    What to do if the CO gas density does not match the set value in Panasonic Accessories?
    • T
      Tanya PenaSep 6, 2025
      The CO gas density may not match the set value if the secondary pressure for the gas regulator is not at the specified value of 0.03 MPa(G) (0.3 kgf/cm2(G), 4.4 psi(G)), the gas tube is blocked, the duct is not securely attached, the fan is not attached properly, or the incubator is operating beside an appliance that generates electromagnetic waves. Attach the duct properly to the 4 points hooks and confirm if the fan is pushed all the way to the motor shaft.
  • J
    Jessica PerrySep 8, 2025
    What causes “Err34: H2O2 Volume” to be activated in Panasonic MCO-230AIC Accessories during H O decontamination cycle?
    • W
      william19Sep 8, 2025
      If the duct is not securely attached, attach it properly to the 4 points hooks. Ensure the fan is attached properly by confirming it is pushed all the way to the motor shaft. Also, check if the H O generator is securely installed by setting the 2 pins of the generator in the holes at the left bottom of the duct. If the total operating time of the H O generator exceeds 5,000 hours, it may be the end-of-life, and you should replace it.
  • C
    Cameron HenrySep 12, 2025
    Why are normal cultures not possible, and the CO gas density is suspect in my Panasonic Accessories incubator?
    • J
      jperkinsSep 12, 2025
      Normal cultures may not be possible, and the CO gas density may be suspect if the ambient air environment around the incubator is not normal, or if there is a source of polluted gas in the vicinity.

Summary

Intended Use and Precautions

Precautions for Safe Operation

Incubator Components

LCD Touch Panel Interface

Explains the main display screen, its fields, and indicators for operation.

Installation Guide

Connecting the CO2 Gas Cylinder

Provides step-by-step instructions for securely connecting the CO2 gas cylinder.

Before Commencing Operation

Initial Cleaning Procedure

Guides the user through cleaning the incubator and its internal components before first use.

For Better Cultivation

Cultivation Precautions

Offers essential advice and precautions for achieving optimal cell cultivation results.

Correct Operation Procedure

Basic Operation on LCD Touch Panel

Basic Parameters and Settings

Alarm Parameters Configuration

UV Lamp Parameters

H2 O2 Decontamination

H2 O2 Decontamination Procedure

Provides comprehensive steps for performing H2O2 decontamination of the incubator chamber.

H2 O2 Reagent Handling Precautions

Lists critical safety precautions for handling the H2O2 reagent and its associated equipment.

Electric Lock (Option)

Gas Auto Changer (Option)

Connecting CO2 Gas Cylinders

Guides on connecting two CO2 gas cylinders to the auto changer system.

STD Gas Auto Calibration Kit (Option)

Routine Maintenance Schedule

Alarms, Safety, and Self-Diagnosis

Troubleshooting Common Issues

Technical Specifications

Performance Characteristics

Panasonic MCO-230AIC Specifications

General IconGeneral
BrandPanasonic
ModelMCO-230AIC
CategoryAccessories
LanguageEnglish

Related product manuals