31
SPECIFICATIONS
Product name
Biomedical Freezer
MDF-237
External dimensions
W905 mm x D687 mm x H881 mm
Internal dimensions
W790 mm x D440 mm x H715 mm
Effective capacity
221 L
Exterior
Painted steel
Interior
Colored aluminum plate
Door
Painted steel
Insulation
Rigid polyurethane foamed-in place
Storage baskets Polyethylene coated steel wire
Access port Diameter 17 mm, Right side and bottom left
Compressor Hermetic type, 200 W
Condenser Wire and tube type
Evaporator Aluminum tube on sheet type
Refrigerant
R-134a
Temperature controller Electronics controller (between -18 and -35
o
C)
Temperature display Digital display (between -50 and +50
o
C)
Temperature sensor Thermister sensor
Temperature alarm Flash of digital indicator and alarm lamp, Buzzer, (Remote alarm)
Accessories 1 set of key, 1 scraper, 3 storage baskets
Weight
60
kg
Battery
For power failure alarm, Nickel-metal-hydride battery, DC 6 V, 1100 mAh
Automatic charge
Optional component
Temperature recorder (MTR-G85C, MTR-4015LH)
Recorder fixing (MDF-S740, MDF-S3040)
2 storage baskets (MDF-13B2), 3 storage baskets (MDF-13B3)
Interface board (MTR-480, MTR-L03),
Data acquisition system (MTR-5000)
Note
:
Design or specifications will be subject to change without notice.
Refer to the updated catalog when ordering an optional component.
The battery for power failure alarm is an article for consumption. It is recommended that the battery will
be replaced about every three years. Contact our sales representative or agent at the time of
replacement of the battery for recycling.