23V1.0/02/12
Technical data
8. Technical data
Version SCM-155-0-02
Analogue version
SCM-155-2-05
CAN version
Inputs Sensor inputs: Two Parker analogue sensor with
sensor ID
Sensor inputs: CAN-bus interface for up to three
CAN BUS Parker sensors with
sensor ID
Measurement accuracy: ≤0.2 % FS ± 1 digit Measurement accuracy: -
Plug connector: 5-pole, Push-Pull Plug connector: 5-pole, M12x1,
SPEEDCON
®
, plug
Polling rate for P-channel: 1 ms Polling rate for P-channel: 1 ms
Interfaces USB device: Online data transmission between instrument and PC via SensoWin software
Measurement transmissions: ACT/MIN/MAX, min. 5 ms
USB standard: 2.0, full speed
Plug connection: Micro-USB connector, shielded, Type B
USB host: Port for USB ash drive (max. 4 GB)
Recommended types: Delock USB 2.0 Nano ash drive, Intenso Micro Line
USB standard: 2.0, full speed,100 mA max.
Plug connection: Micro-USB socket, shielded, type A
Memory Internal memory for measurement data: 1 measurement, about 15,000 records (270,000 readings for
ACT/MIN/MAX)
USB ash drive: 1 GB included
Functions Difference; addition; power;
ACT; MIN; MAX; FS; TEMP display;
Battery status
Start-stop measurement