443B101/443B102 26
Manual 19854 Rev. D
ECN 50523
Command: OFFS
Purpose: Sets the dc offset of the amplifier when in ICP long time time constant mode.
Response: The module responds with a “0” to tell the calling process the command has
been received.
Command String: XYCO2OFFSXX.XXX
Description: X— Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y— Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C02— Module type for the target. Only type “C02” (443B102) can accept this
command since the 443B101 does not have long time constant.
OFFS— Commands the module to set the dc offset.
XX.XXX— 6 character offset voltage. (Range: ASCII printable characters, 0
through 9). Only positive offsets are allowed, up to 20.000 volts.
Response String: “0”
Description: This return indicates that the command was received by the card.
Example: Command String: “02C02OFFS10.361”
Response String: “0”
The dc offset level of the 443B102 in rack 0, and slot 2 will be set to 10.361
volts.