EasyManuals Logo

PeakTech P3730 User Manual

PeakTech P3730
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
28
5.3. Inductance Measurement
To test the inductance, please follow the following procedure:
1. Set the rotary switch to LX measurement mode.
2. If the tested inductance value is unknown, use the maximum
measurement position and decrease the range step by step
until a satisfactory reading is obtained.
3. Insert the test clips into the corresponding Lx input terminals.
4. Use the test clips to clip the inductance to carry out the
testing.
5. The measured value shows on the display.
REMARKS:
* When carrying out measurement at 2 mH, short circuit the test
clip first, then carrying out the measurement. The actual
correct reading is the measure reading minus the short circuit
reading.
* When testing a small value inductance, it is better to use the
small value jack.
* The Meter cannot check the quality of the inductance.
* When inductance measurement has been completed,
disconnect the connection between the testing clips and the
circuit under test and remove the testing clips away from the
input terminals of the Meter.