-30-
MAINTENANCE
Cleaning Procedures (cont’d)
Cleaning Boiler Ring (C)
Clean boiler ring using a non-chlorinated, non-metallic scouring pad
such as Scotch-Brite
®
.
Cleaning Door Gasket (D)
Clean door gasket with Omni-Cleaner Plus
®
or non-chlorinated
detergent and water. Note: If residue is allowed to accumulate, the
seal could be affected and leaks may occur.
Cleaning/Disinfecting Exterior Surfaces
Clean all exterior surfaces with mild detergent and water using a
sponge or cloth. Disinfect exterior surfaces using an iodophor
(Biocide, Biotrol, Inc., N. Salt Lake City, Utah, or equivalent).
! CAUTION: Do Not use any disinfectant on interior stainless
steel surface. Damage to the chamber and/or trays may result.