4.5 Functional
Tidal Volume
Adult bellows 20 to 1600 ml (±10%)
Paediatric bellows 20 to 350 ml (±10%)
At ambient temperature of 20
o
C (+/-10%) and ambient atmosphere of 101.3 kPa (+/-10%).
Minute Volume 0 to 30 L
Rate 4 - 100 bpm
I:E Ratio 1:0.3 - 1:8
Pressure Limit 10 - 80 cmH2O
Fresh Gas Compensation Automatic Tidal Volume adjustment
Modes Off
Standby
Volume Cycle
Pressure Controlled
Spontaneous (includes advanced breathing modes)
Pressure Control 10 - 70 cmH
2O
Inspiratory Flow 2 - 75 L/min
Spontaneous Mode Active Volume and Pressure Alarms,
Advanced Breathing Modes selectable
(see section 4.6)
Electronic PEEP 4 - 30 cmH2O
Oxygen Monitor Fuel Cell type
For full specification, see section 4.15.
Spirometry - Resolution
±1 ml
Ventilator Performance - accuracy of delivered volumes
>300 ml ± 10%
>100 ml <300 ml ± 20%
<100 ml ± 50%.
NOTE
The ventilator is designed for use with Spirometry ON.
Accuracy of delivered volumes with Spirometry OFF may vary from the figures given above.
SPECIFICATION