Sterilisation
Recommended guidelines for sterilisation
CAUTION
To prevent possible damage to components, peak sterilisation
temperatures must not exceed :
54
o
C (130
o
F) for gas (ethylene oxide) or,
134
o
C (275
o
F) for steam autoclave.
Low temperature autoclave is 121
o
C
Do not sterilise the ventilator control unit. The internal
components are not compatible with sterilisation techniques and
may be damaged.
Following sterilisation with ethylene oxide, components must be
quarantined in a well ventilated area to allow dissipation of any
residual gases.
Follow the recommendations given by the steriliser
manufacturer for aeration periods required.
ITEM METHOD
Bellows Gas, liquid, autoclave (20 cycles max.)
Hoses Gas, liquid, autoclave
O rings Gas, liquid, autoclave
Bellows base Gas, liquid, autoclave
Exhalation valve Gas, liquid, pasteurise, low temperature
assembly autoclave
Control unit Do not sterilise - remove patient block
Patient Block Autoclave
Bellows canister Liquid, autoclave
Oxygen monitor (including sensor) - see section 6.4
NOTE
1. Liquid method indicates the use of a high level
disinfectant.
2. Examples of suitable high level disinfection liquid agents
are: Nu-Cidex, Sporicidin, and Sonacide.
3. The exhalation diaphragm valve must be removed,
cleaned and sterilised separately.
APPENDIX
87