4.1 Application Ventilation for use in anaesthesia.
4.2 Internal Compliance
Adult bellows 3 ml/cmH
2O (nominal)
Paediatric bellows 2 ml/cmH
2O (nominal)
4.3 Physical
Size (mm)
- control unit only 290 wide x 300 deep x 185 high
- with adult bellows 290 wide x 300 deep x 385 high
Screen Size 210 mm (8.4") TFT
Weight - control unit only 7.6 kg
- with adult bellows 9 kg
Bellows
Adult (Latex free): 20 ml - 1600 ml
Paediatric : 20 - 350 ml
(Note - latex free paediatric available as option)
Power 90 - 264 VAC, 47 - 63 Hz
Battery Back-up: 30 minutes (assumes fully charged battery)
Drive Gas Oxygen or Air
(dry, and oil free) at 45 to 100 psi (310 to 689 kPa).
4.4 Alarms
Alarm Mute 30 or 120 seconds (see 3.11)
Low Drive Gas Pressure Less than 235 kPa (35 psi)
High Continuous Airway Pressure Above 30 cmH
2O at start of cycle
Low Pressure 4 - 14 cmH2O PEEP referenced
Low Tidal Volume 50% of Volume set (Spirometry)
Incorrect Rate or Ratio
Mains Failure 30 minutes (nominal) Battery Backup
Low Battery 5 minutes Use
Ventilator Inoperative Internal or Battery Failure
Alarms - User Set
Vt (Tidal)
Min 0 - 1600 ml
Max 20 - 1600 ml
Vm (Minute)
Min 0 - 10 L
Max 0 - 30 L
Apnoea Adjustable Re-set Pressure (PEEP referenced)
Low and High O2 Concentration 18% - 105%
High Airway Pressure 10 - 80 cmH2O adjustable
4. SPECIFICATION