Operating the Anaesthesia System
6.12.4 Alarm message
When an alarm occurs, a message appears n the alarm area
The alarm area background colour matches the alarm level
Hgh level alarms red
Medum level alarms yellow
Low level alarms yellow
For physologcal alarms, the symbol () dentfes alarm level
Hgh level alarm
Medum level alarm
Low level alarm
6.12.5 Clinical alarms
Message Prorty ause Acton
APNEA Hgh
Breathng or ventlaton has
stopped
heck patent’s spontaneous breathng
ablty
heck for blockages n the breathng
crcut
ontnuous Pressure Hgh Hgh
Arway pressure greater than
(PEEP +15) cmH
O for 15 seconds
heck for blockages n the breathng
crcut
PRESSURE HIH Hgh
Ppeak s hgher than the Paw hgh
alarm lmt settng
Decrease tdal volume or ncrease Paw
hgh alarm lmt value
heck for blockages n the patent crcut
FO < 18% Hgh FO s lower than 18%
Recalbrate the O
2
cell
FHAL HIH Hgh FAA s greater than alarm lmt
Set the alarm lmts approprately
heck agent settng
FENF HIH Hgh
FISO HIH Hgh
FSEV HIH Hgh
FDES HIH Hgh
EtHAL HIH Hgh EtAA s greater than alarm lmt
Set the alarm lmts approprately
heck agent settng
EtENF HIH Hgh
EtISO HIH Hgh
EtSEV HIH Hgh
EtDES HIH Hgh
MV LOW Hgh
MV s lower than the low alarm
lmt settng
Increase tdal volume or breath rate, or
decrease low alarm lmt
MV HIH Hgh
MV s hgher than the hgh alarm
lmt settng
Decrease tdal volume or breath rate, or
ncrease hgh alarm lmt
PRESSURE < -10 cmH
2
O Hgh Paw s less than -10 cmHO
heck for blockages n the breathng
crcut
Increase fresh gas flow
heck that there s hgh flow gas flowng
through the ASS If yes, check the
negatve pressure relef valve on the
recever