Operating the Anaesthesia System
6.12.4 Alarm message
When an alarm occurs, a message appears n the alarm area
The alarm area background colour matches the alarm level
Hgh level alarms red
Medum level alarms yellow
Low level alarms yellow
For physologcal alarms, the symbol () dentfes alarm level
Hgh level alarm 
Medum level alarm 
Low level alarm 
6.12.5 Clinical alarms
Message Prorty ause Acton
APNEA Hgh
Breathng or ventlaton has
stopped
heck patent’s spontaneous breathng
ablty
heck for blockages n the breathng
crcut
ontnuous Pressure Hgh Hgh
Arway pressure greater than
(PEEP +15) cmH
2
O for 15 seconds
heck for blockages n the breathng
crcut
PRESSURE HIH Hgh
Ppeak s hgher than the Paw hgh
alarm lmt settng
Decrease tdal volume or ncrease Paw
hgh alarm lmt value
heck for blockages n the patent crcut
FO < 18% Hgh FO s lower than 18%
Recalbrate the O
2
cell
FHAL HIH Hgh FAA s greater than alarm lmt
Set the alarm lmts approprately
heck agent settng
FENF HIH Hgh
FISO HIH Hgh
FSEV HIH Hgh
FDES HIH Hgh
EtHAL HIH Hgh EtAA s greater than alarm lmt
Set the alarm lmts approprately
heck agent settng
EtENF HIH Hgh
EtISO HIH Hgh
EtSEV HIH Hgh
EtDES HIH Hgh
MV LOW Hgh
MV s lower than the low alarm
lmt settng
Increase tdal volume or breath rate, or
decrease low alarm lmt
MV HIH Hgh
MV s hgher than the hgh alarm
lmt settng
Decrease tdal volume or breath rate, or
ncrease hgh alarm lmt
PRESSURE < -10 cmH
2
O Hgh Paw s less than -10 cmHO
heck for blockages n the breathng
crcut
Increase fresh gas flow
heck that there s hgh flow gas flowng
through the ASS If yes, check the
negatve pressure relef valve on the
recever