Prma 465 Anaesthetc Machne 67 User Manual
Operating the Anaesthesia System
Message Prorty ause Acton
EtHAL LOW
Medum EtAA s less than alarm lmt
heck the fll level of the vaporzers
Set alarm lmt approprately
heck agent settng
EtENF LOW
EtISO LOW
EtSEV LOW
EtDES LOW
FNO LOW Medum FNO s less than alarm lmt
Set the requred alarm lmts
heck the NO settng
Mxed Agent and MA > =3 Medum
Two dfferent agents are detected
and the MA calculaton s greater
than or equal to 3
Make sure only one vaporzer s on
Wat approxmately two mnutes for the
frst agent to wash out of the system
Mxed Agent and MA < 3 Low
Two dfferent agents are detected
and the MA calculaton s less
than 3
Make sure only one vaporzer s on Wat
approxmately two mnutes for the frst
agent to wash out of the system
6.12.6 Technical alarms
Message Prorty ause Acton
O SUPPLY DOWN Hgh
O ppelne pressure s less than
280 kPa
heck the O ppelne and cylnder are
properly connected
BATTERY DISHARED Hgh
Remanng battery power s
between fve and ffteen mnutes
Plug n the power cable
heck that the system crcut breaker s
on
EM FAILURE Hgh Electronc gas mx hardware error
Swtch to alternate O control
Restart anaesthetc workstaton
ontact your servce personnel
EM communcaton falure Hgh Hardware error
Restart anaesthetc workstaton
ontact your servce personnel
BDU communcaton falure Hgh
Breath delvery unt hardware
error
Restart anaesthetc workstaton
ontact your servce personnel
as Module Software Error Hgh as Module Software Error ontact your servce personnel
as Module Hardware Error Hgh as Module Hardware Error ontact your servce personnel
as Module Motor Overspeed Hgh as Module Motor Overspeed ontact your servce personnel
as Module Factory calb lost Hgh as Module Factory calb lost ontact your servce personnel
as Module Replace Adapter Hgh as Module Replace Adapter ontact your servce personnel
as Module No Adapter Hgh as Module No Adapter ontact your servce personnel
O Unspecfed Accuracy Hgh O Unspecfed Accuracy ontact your servce personnel
NO Unspecfed Accuracy Hgh NO Unspecfed Accuracy ontact your servce personnel
AA Unspecfed Accuracy Hgh AA Unspecfed Accuracy ontact your servce personnel
as Module Temp Out Of
Range
Hgh as Module Temp Out Of Range ontact your servce personnel
Pressure Out Of Range Hgh Pressure Out Of Range ontact your servce personnel
AA ID Unrelable Hgh AA ID Unrelable ontact your servce personnel
BATTERY LOW Hgh
Remanng battery power s
between ten and thrty mnutes
Plug n the power cable
heck that the system crcut breaker s
on