Prma 465 Anaesthetc Machne 3 User Manual
20 Portable RF (rado frequency) communcatons
equpment (ncludng perpherals such as antenna
cables and external antennas) should be used no
closer than 30 cm (12 nches) to any part of ths devce,
ncludng cables specfed by the manufacturer
Otherwse, degradaton of the performance of ths
equpment could result
21 Addtonal equpment placed on the top shelf must be
securely attached
22 The followng condtons must apply before the machne
s transported
a) Detachable parts removed
b) The drawers must be empty
c) All gas cylnders must be removed
23 MRI compatblty - Prma 465 models are not MRI
compatble
24 To prevent patent nury n the event of total anaesthesa
system falure, an alternatve means of ventlaton must
be avalable whenever the devce s n use
25 Obstructon of the breathng system can restrct or
stop gas flow to the patent, and can cause nury or
death Make sure that there are no obstructons n the
breathng system
26 Keep very small components/parts/plugs away from the
breathng system
27 Use non-conductve breathng system hoses
28 Do not touch any electrcal devce connector at the same
tme as the patent, when the machne s connected to
the mans supply
29 Ths machne s not sutable for use n oxygen-rch
envronment
30 Ths devce must not be altered or modfed n any way
wthout the wrtten permsson of Penlon lmted
31 Make sure that the absorber canster sensor s clean and
unobstructed by cables or tubng etc, at all tmes
32 ommuncaton ports should not be subected to
voltages greater than 7 V D Damage to nternal
crcutry may occur
33 In the event of a mans falure or operaton of the mans
crcut breaker, the auxlary mans outlet sockets wll
not functon Make sure that any equpment connected to
the auxlary mans outlet sockets are protected aganst
mans falure
Warnings and Cautions
34 Ventlatng neonates (nfants)
a) Use only Pressure ontrol Ventlaton or Volume
ontrol ventlaton for mandatory ventlaton of
neonates
b) Always set a low target pressure frst and then
ncrease to the requred level
c) Use breathng crcuts and flters specfcally desgned
for paedatrc/neonatal use
d) Do not ventlate pre-term babes wth brth weghts
below 24 kg
35 Patent montorng
a) Always use an anaesthetc agent montor that
comples wth the latest relevant standard
b) Always montor oxygen concentratons
c) Always check that the montors are functonng
correctly before startng the clncal procedure
d) An external anaesthetc gas montorng devce,
complyng wth ISO/IE 80601-2-55 must be
connected to the breathng crcut Use the
manufacturer’s nstructons to correctly connect the
devce
36 When connectng other equpment to the auxlary sockets a
medcal electrcal system s created and the whole system
should be re-evaluated to the requrements of BS EN
60601-1
37 as analyser (NO, O, O) measurements are ntended
only to assst n patent assessment These devces must
be used n conuncton wth other patent physologcal
ndcatons and n the assessment of the overall patent
condton
38 When any accessory s nstalled on the shelf or attached
to an arm mount, you must check the stablty of the
anaesthetc machne before use
39 When the machne s n operaton, the alarm message
feld must be vsble to the operator at all tmes
40 Sdestream anaesthetc gas montorng module (f
ftted) heck the water trap for damage (cracks, etc) and
replace f damaged
To mantan adequate levels of hygene, the water trap must
be replaced every four weeks
Note that the sample tubng s a sngle-use component