11 BATTERY CHARGING
The battery in the instrument can be re-charged by the Redpost RPC-80
printer/charger or the Redpost RPC-50 battery charger. You can connect the
charger to either socket. If you connect to the charger before recording a file you
see the complete charging information on the screen.
*
Warning
Do not leave the battery on charge continuously for more than 24
hours.
11.1 BATTERY INFORMATION
All the other display screens show the battery condition using just the small
battery image from the complete diagram. This shows when the battery is being
charged by an arrow head that moves up the battery image. The speed of
movement shows the charge rate. The dotted line shows the minimum level of
charge needed to start recording.
11.2 BATTERY TEMPERATURE
The charge rate is controlled by several factors including battery temperature.
The battery will not charge if it is too hot and in that case the arrow head in the
battery diagram will be stationary. Charging will resume as soon as the battery
has cooled.
*
The illustration shows the English language version. The text changes according to the language selected for the
display.
Instrument type and serial number
Battery
condition*
Software
version
Charge
rate*
Moves faster for
higher charge rate
Charge above here
before recording
Indicates that charger
Level of black filling
shows battery