- 17 -
10) Tap “Start” button on the next screen (Photo-19 Left). And the Updating starts (Photo-19 Middle).
11) After the Updating finishes, message “Complete” appears on the next screen (Photo-19 Right).
Note: If the Updating has failed, message “Complete” will not appear. In this case tap “Quit” button and repeat
the procedures from 6).
12) Turn OFF the EPK-i.
6.5.4 Pump air flow data restore
Restore the pump air flow data with the data that has been recorded in the section “6.5.1 Pump air flow data
backup”
<Procedures>
1) Restore the pump air flow data to the new Process Board (H700) with the backup data with the
procedures 2) - 9) of “6.2.2 Pump air flow adjustment”.
2) If the backup has failed in the section “6.5.1 Pump air flow data backup”, perform “6.2.2 Pump air flow
adjustment”.
Tap ”Start” button
Tap “Quit” if necessary
Message “Complete”
Photo-19 Touch panel operation -4