EasyManua.ls Logo

Pepperl+Fuchs KCD2-UT2-1 - Page 25

Pepperl+Fuchs KCD2-UT2-1
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Functional Safety KCD2-UT2-(Ex)1, HiC2081
Operation
2018-02
25
5.3.2 Resistance Thermometer Input (RTD)
Figure 5.6 Proof test set-up for HiC2081 with resistance thermometer input (RTD)
Proof Test Procedure for Resistance Thermometer Input
1. Connect a RTD simulator, depending on the application.
for 4-wire connection: terminals 1, 2, 5 and 4
for 3-wire connection: terminals 1, 5 and 4
for 2-wire connections: terminals 1 and 4
2. Connect the digital multimeter to the terminals 11 and 14.
3. Set the RTD simulator sequentially to the temperature values
representing 4 mA, 12 mA, 20 mA at the output.
4. Measure the output current.
The proof test is passed if the following values are measured at the output
1
:
for the 4 mA application: 3,7 mA to 4,3 mA
for the 12 mA application: 11,7 mA to 12,3 mA
for the 20 mA application: 19.7 mA to 20.3 mA
5. Apply a short circuit between terminal 1 and 4. Do not remove the RTD simulator. Check if
a short circuit of the cold junction compensation is detected.
The red LED is flashing. The output behavior in the event of a fault depends on the
device configuration.
6. Remove the RTD simulator. Check if a lead breakage is detected.
The red LED is flashing. The output behavior in the event of a fault depends on the
device configuration.
7. Set the device back to the original settings after the test.
11
14
1
2
5
4
HiC2081
Termination Board
Zone 0, 1, 2
Fault
+
-
+
SL2
5a
1a
-
1b
5b
SL1
8a
7a
Multimeter
(mA)
T
T
T
Supply +
Supply -
24 V DC
Power
supply
I supply
Supply
Bus
1
Additionally the loop diagnosis shall be tested to prove that the fault signalling via the current output is working
correctly. The output current in the event of a failure depends on the device configuration. Please record this
configuration and the resulting expected fault signalling current in the test report. Example: if downscale is
configured, 2.0 mA ±1 % must be measured in the event of a failure. The red LED must be flashing.

Related product manuals