EasyManua.ls Logo

PERSEE T6 U - Page 18

PERSEE T6 U
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Setting the working wavelength
In the case of photometric measurement (Fig.3-1), pressGOTOλ to set wavelength, as shown in
Fig.3-9.
(Fig. 3-9)
At the bottom of this page, enter the wavelength by pressing the numeric keys (0~9) and decimal point
【.. Press RETURN key to clear the mistakes, and press CE to clear step by step, then press
ENTER key to confirm . If nothing is to be entered, press RETURN key to quit this page
directly. The range is from 190 to 1100. If the entered figure exceeds the range, the system will delete what
has been entered automatically and you should start a new entry again.
Auto Zero
In the case of photometric measurement (Fig.3-1), press ZERO to zero the Abs (or 100%T) at
current wavelength. Put the blank sample into the sample compartment before correction.
Measurement
In the case of photometric measurement (Fig.3-1), press START to gain access to the measurement
menu, then press START again to measure the samples at current wavelength in the auto cell changer
or the fixed cell holder, as shown in Fig.3-10.
(Fig. 3-10)
500.0 nm
The current wavelength
1.000 Abs
The actual measured value of current cell
NO.1-1
The first measurement value of No.1 Cell
Abs
Current measurement mode
K*Abs
Factor K multiply by the current Abs
Five rows of data can be displayed on the screen, the others can be displayed by pressing
【▲】
and
【▼】
.
Printout