EasyManua.ls Logo

PERSEE T6 U - Page 17

PERSEE T6 U
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
blank value. If you select yes for the cell blank, the test is necessary. Put blank samples in each cell before
test, the system will record the measurement values of each cell automatically.
If you select yes for both reagent blank and cell blank, you can perform cell blank for each cell and
take the sample in No.1 cell as the blank to correct the measured values of other cells.
Blank correction is a kind of data processing. The following table shows the calculation.
Cell
number
Reagent blank and
Cell blank are
NO
Only the
reagent
blank
YES
Only the
cell blank
YES
Both blank
correction
YES
The
output
data
NO. n
sample
cell
(n=2---8)
Abs
An
An - a1
An - an
(An - an) -
( A1 - a1)
Transmittance
Tn
Tn / t1
×
100(%)
Tn / tn
×
100(%)
Tn * t1 / tn *
T1
×
100(%)
NO. 1
sample
cell (as
the
reagent
sample
cell)
Abs
A1
A1
A1 - a1
A1 - a1
Transmittance
T1
T1
T1 / t1
×
100(%)
T1 / t1
×
100(%)
An
The Abs of NO. n sample cell without
correction
Tn
The transmittance of NO. n sample cell
without correction
an
The blank value of NO. n sample cell (Abs)
tn
The blank value of NO. n sample cell
(transmittance)
Move Cell
Press 【▲】 and 【▼】 to move the cursor to the required option, then press ENTER. The sample
cell will move forward one cell by pressingENTEReach time. When it moves to the maximum cell
number, next it will return to No.1 cell automatically. If the user selects fixed cell, the option is
inoperative.
Factor K
Input the required magnified multiple times of the measurement value.
Press 【▲】 and 【▼】 to move the cursor to the required option, then press ENTER to set
factor K. At the bottom of this page, enter the factor K by pressing the numeric keys (0~9), decimal point
【.and negative sign-. PressRETURNkey to clear the mistakes, and pressCE to clear step
by step, then pressENTERkey to confirm . If nothing is to be entered, press RETURN key to quit
this page directly. The range is from -99 to 99. If the entered figure exceeds the range, the system will
delete what has been entered automatically and you should start a new entry again.
Cell Init.
If the user moves the cell during measurement, cell init. should be performed.
Press 【▲】 and 【▼】 to move the cursor to the required option, then press ENTER to perform
cell init. It will return to No.1 cell.