Cleaning CritterCuffs: Clean cuffs with a damp cloth or
hand wash in warm water with mild detergent or soap; do not
allow water to get into the cuff tube. Pat cuff dry initially on a
towel, and air dry until fully dry and ready for use.
Cleaning petMAP: Clean with a cloth dampened with warm
water or a mild bleach solution. Do not use alcohol or
cleaning solutions which contain alcohol. Do not immerse in
water or other liquids.
Troubleshooting
NOTE: At the first indication of any petMAP malfunction,
the first attempt at resolving the problem should be to
replace the batteries with known fresh batteries.
Difficulty obtaining consistent BP readings
(greater than +/- 15 mmHg from reading to reading)
1. Animal is moving. Calm animal until they are relaxed so
determinations can be made during periods of
non-movement.
2. Wrong size cuff used; use a cuff sized so that the index
line is in the “optimum” region when snugly fitted to the
limb.
3. Cuff not properly snug, i.e., not tight enough on
initial placement. Tighten cuff on the limb.
4. Cuff is not in correct location on forearm, hind foot, or
tail. Reposition cuff or move to another approved site
and set petMAP using multi-function button.
5. Check LoBat light. Replace batteries if indicated.
6. Zero not set properly on power up. Make sure there is
no pressure in the system when the petMAP is first
powered up by opening the deflate trigger first and/or
removing the cuff connector from the unit to assure
atmospheric pressure, before turning Power ON.
Maintenance