7.3 Changing lubricant
WARNING
Health hazard and environmental damage from toxic contaminated lubricant
Toxic process media can cause lubricant contamination. When changing the lubricant, there is a
health hazard due to contact with poisonous substances. Illegal disposal of toxic substances causes
environmental damage.
Wear suitable personal protective equipment when handling these media.
Dispose of the lubricant according to locally applicable regulations.
CAUTION
Scalding from hot lubricant
Danger of scalding when draining lubricant if it comes into contact with the skin.
Wear protective equipment.
Use a suitable collection receptacle.
Pfeiffer Vacuum recommends determining the precise service life of the lubricant in
the first operating year.
The usable life may deviate from the reference value specified depending on thermic and
chemical loads, or due to penetrating process media in gear and bearing chambers.
Safety data sheets
You can obtain the safety data sheets for lubricants from Pfeiffer Vacuum on request, or
from the Pfeiffer Vacuum Download Center.
7.3.1 Draining lubricant
Required tools
Ring spanner, WAF 17 mm
Calibrated torque wrench (tightening factor ≤ 2.5)
9
8
3
2
1
4
6
7
5
Fig. 10: Draining lubricant
1 O-ring 6 O-ring
2 Filler screw 7 Collection receptacle
3 Filler screw 8 Drain screw
4 O-ring 9 O-ring
5 Drain screw
Maintenance
34/54

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pfeiffer Vacuum OKTA 500 and is the answer not in the manual?

Pfeiffer Vacuum OKTA 500 Specifications

General IconGeneral
BrandPfeiffer Vacuum
ModelOKTA 500
CategoryWater Pump
LanguageEnglish

Related product manuals