Notice
Please add the following NOTICE to the original manual, SM9910171.
P.15
Fan motor lifespan check:
<Reset of accumulation time>
Step into F06. Input
419
and press SET key to reset
non-volatile memory to ’00.0’. The display reverts to chamber
temperature only.
P.19 F32
<Reset>
Step into F06. Input
419
and press SET key to reset non-volatile
memory to ’00.0’. The display reverts to chamber temperature
only.
Corrected as follow
P.15
Fan motor lifespan check:
<Reset of accumulation time>
Step into F06. Input
410
and press SET key to reset
non-volatile memory to ’00.0’. The display reverts to chamber
temperature only.
P.19 F32
<Reset>
Step into F06. Input
410
and press SET key to reset non-volatile
memory to ’00.0’. The display reverts to chamber temperature
only.
Panasonic Healthcare Co., Ltd.
CORRECTION
SERVICE
PRODUCT IMPROVEMET
PRODUCTION CHANGE
ADD INFORMATION
FOR RoHs
FILE NO.
Category: ULTRA-LOW TEMP. FREEZER Date: Jun.8, 2015
Model: MDF-C8V1 Destination: ALL
Ref No: SM9910171 Issued No. E230
REFERENCE NO. SM9910171
SM9910171SM9910171
SM9910171-
--
-05
0505
05
NOTICE
This is the correction for the incorrect descriptions about "Reset of accumulation time" of
fan motor in the page15 and 19 of Service Manual, SM9910171.