<Distribution data>
Temperature of the cycle in each area (SV=-70℃、air temperature)
Unit:℃
Maximum Minimum
Middle of
cycle
Differential
Left back
-67.1 -71.5 -69.3 ±2.2
Right back
-67.1 -71.3 -69.2 ±2.1
Right front
-67.4 -71.5 -69.5 ±2.1
Left front
-67.3 -71.7 -69.5 ±2.2
Center
-67.4 -70.4 -68.9 ±1.5
Left back
-67.9 -71.4 -69.7 ±1.8
Right back
-67.6 -71.4 -69.5 ±1.9
Right front
-67.6 -71.4 -69.5 ±1.9
Left front
-67.9 -71.4 -69.7 ±1.8
Center
-68.1 -70.7 -69.4 ±1.3
Left back
-67.5 -70.4 -69.0 ±1.5
Right back
-67.3 -70.6 -69.0 ±1.7
Right front
-67.2 -70.8 -69.0 ±1.8
Left front
-67.5 -70.9 -69.2 ±1.7
Center
-67.6 -70.3 -69.0 ±1.3
- - -69.3 -
Unit:℃
Maximum Minimum
Middle of
cycle
Differential
Left back
-67.1 -71.2 -69.2 ±2.1
Right back
-67.1 -71.2 -69.2 ±2.1
Right front
-67.4 -71.3 -69.4 ±2.0
Left front
-67.4 -71.5 -69.5 ±2.1
Center
-67.5 -70.3 -68.9 ±1.4
Left back
-68.0 -71.2 -69.6 ±1.6
Right back
-67.7 -71.2 -69.5 ±1.8
Right front
-67.7 -71.2 -69.5 ±1.8
Left front
-67.9 -71.2 -69.6 ±1.7
Center
-68.1 -70.6 -69.4 ±1.3
Left back
-67.4 -70.3 -68.9 ±1.5
Right back
-67.1 -70.3 -68.7 ±1.6
Right front
-67.0 -70.7 -68.9 ±1.9
Left front
-67.4 -70.7 -69.1 ±1.7
Center
-67.5 -70.2 -68.9 ±1.4
- - -69.2 -
<Amount of power consumption>
Amount of power consumption when driving at cycle
(SV=-70℃
1φ220V
Unit:kWh/day
50Hz 60Hz
AT23℃
3.67
AT30℃ 4.04
Average
Note:This data does not represent a guarantee of product performance.
Note:This data does not represent a guarantee
of product performance.
Upper area
Middle area
Bottom area
Ambient temperature 30℃
220V 60Hz
Upper area
Middle area
Bottom area
Average
Ambient temperature 23℃
220V 60Hz
Page 20 of 27