EasyManua.ls Logo

Phcbi MDF-U443 - Function Mode; Setting of Alarm Delay Time

Phcbi MDF-U443
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
FUNCTION MODE
This unit has the following function mode.
Indication Mode Settable range
F01 Setting of high temperature alarm
Between +5
o
C and 20
o
C higher than the
chamber set temperature (1
o
C gradient)
F02 Setting of low temperature alarm
Between -5
o
C and -20
o
C lower than the
chamber set temperature (1
o
C gradient)
F04 Setting of delay time of door alarm
Between 1 and 15 minutes
(1 minute gradient)
F05
Setting of compressor delay time (for low
stage side only)
Between 3 and 15 minutes
(1 minute gradient)
F25 Setting of alarm resume time
000 or between 10 and 60 minutes
(10-minute gradient)
F50 Setting of alarm delay time
Between 0 and 15 minutes
(1 minute gradient)
SETTING OF ALARM DELAY TIME
The delay time of the buzzer and remote alarm for high and low temperature alarm can be set between 0
and 15 minutes. The procedure in table 3 shows the sequence to set the delay time to 10 minutes.
The delay time is set to 15 minutes at the factory.
Table 3 Setting procedure for alarm delay time (change from 15 minutes to 10 minutes)
Description of operation Key operated Indication after operation
1 ----
The current chamber temperature is
displayed.
2
Press numerical value shift key for
5 seconds.
The first digit is flashed.
3
Scroll the figure to F50 by using
digit shift key and numerical value
shift key
When pressed, the figure of settable
digit changes.
When pressed, the changeable digit
moves.
4 Press set key. SET
The current delay time is displayed.
The first digit is flashed.
5
Set the figure to 010 with the
numerical value shift key.
When pressed, the figure of the first
digit changes.
6 Press set key. SET
The delay time is memorized and
the current chamber temperature is
displayed.
The alarm delay time set mode returns to the temperature display mode automatically when 90
seconds has passed without any key operation.