6CP100/50-10
3-4
CONTROLS AND INDICATORS
High Voltage Transformer (cont’d)
5. Return Binding Post (RTN)
Low potential side of test specimen connects here. This connection routes current through the Current
Meter circuit (metered connection point).
6. TX1 Connector
Signal / Metering cable connects from here to TX1 on control box.
7. 50 kV TAP
For 0 50 kV output at 0 200 mA the Measurement Divider switch on the control panel must be set
to 50 kV/200 mA position.
8. 100 kV Tap
For 0 100 kV output at 0 100 mA - the Measurement Divider switch on the control panel must be
set to 100 kV / 10 mA.