6CP200/100-10
3-2
CONTROLS AND INDICATORS
Control Panel (Cont’d)
8. Emergency Off switch. Press to immediately shut off all power.
9. High Voltage On Switch and Indicator Lamp. Turns on output voltage when Ready indicator is
illuminated.
10. High Voltage Off Switch and Ready Indicator Lamp. Turns HV off, indicator shows when all
conditions are met to turn H.V. ON. (External Interlock closed, Reset Lamp off, Emergency Off Button
pulled up, Voltage Control at Zero).
11. Voltage Control. Adjusts Output Voltage and must be set at zero to turn High Voltage on.
12. Currentmeter Range Selector.
13. Output Currentmeter.
14. Output Voltmeter.
NOTE: 200kV voltmeter range is calibrated at 80% of range. 100kV range may not be accurate above
100% of range.
15. SX1 External Interlock. Provides for user connection of external interlock or auxiliary safety control
device such as Emergency Off Switch, Gate Switch, Footswitch, Dead Man Switch, etc.
16. Output to Transformer. High Voltage transformer power connection.
17. Main Input. Main input cable connects here.
18. Ground Terminal.
19. TX1. Signal/Metering cable between HV Cylinder and controls connects here.
20. DX1. Signal/Metering cable between HV Divider and controls connects here.
21. Timer Press START/RESET to start the timer after test voltage has been reached. Press
START/RESET again to reset timer to the original setting. Press HV OFF BYPASS to allow high
voltage to remain on after the timer has expired. Test time can be set using the buttons on the face of
the timer.