6 Patient Alarms and INOPs
77
CHECK TI CONFIG Monitor If this INOP appears, the monitor can be used normally, except
for the Trace Interpretation feature. Contact your service
personnel.
Check Touch Input Monitor Perform a visual and functional check of the touch input device.
Contact your service personnel.
COINCIDENCE Monitor One or more fetal/maternal heart rates/pulse rates persistently
coincide with each other (see “Cross-Channel Verification” on
page 97).
CUFF NOT DEFLATED
Numeric is replaced by a -?-
INOP tone.
During this INOP, alarms
cannot be paused or switched
off.
Noninvasive
blood pressure
Remove the cuff from the patient. Make sure that the tubing is
not kinked or twisted. Try restarting the measurement. You can
silence the INOP, but the INOP message remains visible until
the next measurement is started or the
Stop All SmartKey is
selected.
DFHR1 EQUIP MALF
DFHR2 EQUIP MALF
DFHR3 EQUIP MALF
INOP tone
DECG There is a problem with the DECG hardware. Contact your
service personnel.
DFHR1 LEADS OFF
DFHR2 LEADS OFF
DFHR3 LEADS OFF
Numeric is replaced by a -?-
INOP tone.
DECG One or more DECG lead is not attached. Make sure that all
required leads are attached, and no electrodes have been
displaced. Check all connections are sound, and that the legplate
attachment electrode is properly attached. If the INOP persists,
try using another adapter cable, or legplate attachment electrode.
If the INOP still persists, contact your service personnel.
DFHR1 SIGNAL LOSS
DFHR2 SIGNAL LOSS
DFHR3 SIGNAL LOSS
DECG The input signal quality is not sufficient to process the
measurement. Reapply the fetal scalp electrode.
DFHR1 UNPLUGGED
DFHR2 UNPLUGGED
DFHR3 UNPLUGGED
INOP tone
DECG Reconnect the DECG transducer to the monitor. Check all
connections are sound.
MECG EQUIP MALF
INOP tone
MECG There is a problem with the MECG hardware. Contact your
service personnel.
MECG LEADS OFF
Numeric is replaced by a -?-
INOP tone.
MECG One or more MECG lead is not attached. Make sure that all
required leads are attached, and no electrodes have been
displaced. Check all connections are sound. If the INOP
persists, try using another adapter cable. If the INOP still
persists, contact your service personnel.
MECG UNPLUGGED
INOP tone
MECG Reconnect the MECG transducer to the monitor. Check all
connections are sound.
FetRec EQUIP MALF
INOP tone
Recorder There is a problem with the fetal recorder hardware. Contact
your service personnel.