6 Patient Alarms and INOPs
78
FetRec CHK CONF.
INOP tone
Recorder
Recorder Speed and/or Scale Type settings are set to
"Unknown" and need to be set to the correct values in
Configuration Mode before the recorder can be operated.
FHR1 EQUIP MALF
FHR2 EQUIP MALF
FHR3 EQUIP MALF
INOP tone
FHR
(ultrasound)
There is a problem with the FHR hardware. Contact your
service personnel.
FHR1 SIGNAL LOSS
FHR2 SIGNAL LOSS
FHR3 SIGNAL LOSS
FHR
(ultrasound)
The input signal quality is not sufficient to process the
measurement. Adjust the position of the transducer to obtain a
better signal.
FHR1 UNPLUGGED
FHR2 UNPLUGGED
FHR3 UNPLUGGED
INOP tone
FHR
(ultrasound)
Reconnect the FHR transducer to the monitor. Check all
connections are sound.
Internal.Comm.Malf
INOP tone
Monitor There is a problem with I2C Bus communication in the monitor.
Contact your service personnel.
IUP EQUIP MALF
INOP tone
IUP There is a problem with the IUP hardware. Contact your service
personnel.
IUP UNPLUGGED
INOP tone
IUP Reconnect the IUP transducer to the monitor. Check all
connections are sound.
NBP CUFF OVERPRESS
Numeric replaced by a -?-
INOP tone.
During this INOP, alarms
cannot be paused or switched
off.
Noninvasive
blood pressure
The cuff pressure exceeds the overpressure safety limits.
Remove the cuff from the patient. Make sure that the tubing is
not kinked or twisted and that the correct patient category is
selected. Try restarting the measurement.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
Stop All
SmartKey is selected.
NBP EQUIP MALF
Numeric is replaced by a -?-
INOP tone.
Noninvasive
blood pressure
Remove the cuff from the patient. The noninvasive blood
pressure hardware is faulty. Contact your service personnel.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
Stop All
SmartKey is selected.
NBP INTERRUPTED
Numeric is replaced by a -?-
INOP tone.
Noninvasive
blood pressure
Check the tubing and cuff for leakages or kinks. Check that you
are using the correct cuff size and placement, and that the
correct patient category is selected. Try restarting the
measurement.
If the INOP occurs repeatedly, contact your service personnel.
You can silence this INOP, but the INOP message remains
visible until the next measurement is started or the
Stop All
SmartKey is selected.
This INOP arises when the measurement needed longer than
the maximum time for inflation, deflation or the total
measurement.
INOP Message, Indication Source What to do