25
English
Single electric breast pump Double electric breast pump
Battery Rechargeable battery: 3.6 V 2600 mAh (non-replaceable)
Power port Micro-USB
Operating conditions Temperature: 5 °C to 40 °C (41 °F to 104 °F)
Relative humidity: 15 to 90 % (non-condensing)
Storage and transportation
conditions
Temperature: -20 °C to 45 °C (-4 °F to 113 °F)
Relative humidity: up to 90 % (non-condensing)
Operating and storage
pressure
700-1060 hPa (<3000 m altitude)
Net weight 230 g 310 g
External dimensions Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
Device classication Applicable classications:
Degree of protection against electric shock:
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specied as medical electrical system according
to IEC 60601-1.
Materials ABS, silicone (motor unit)
Polypropylene, silicone (other parts)
Service life 500 hours
Mode of operation Continuous operation Non-continuous operation
55 min on, 180 min o at 22 °C
(72 °F) (duty cycle 23 %)
20 min on, 90 min o at 40 °C
(104 °F) (duty cycle 18 %)
Ingress protection IP22 (motor unit), IP20 (adapter)
Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic compatibility
(EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying tables. Only use this Philips
Avent single/double electric breast pump Premium Plus in combination with the adapter supplied.
Emissions Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The device does not use RF energy for its intended
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
English