Software Update Procedures
1MAN0158-H11 04-2005 CSIP Level 1 CardioMD Service Manual 7-13
Procedure Details
6. Insert the software installation CD-ROM
in the acquisition PC’s CD-ROM drive.
To program the collimator ID, you use the
terminal emulator program ttermpro. Proceed
as follows:
7. In the Windows Explorer, select the folder
EDC CardioMD programming for
Module rev. 9CSY0900-B02 or
later\TTERMPRO
on the CD-ROM.
8. In this folder, double-click the icon
ttermpro.exe to start the ttermpro pro-
gram.
9. Press [Enter] to check if the terminal
prompt
-> is present. If the prompt does
not appear, check the serial port setup (se-
lect Serial port from the Setup menu):
Correct COM port selected
Baudrate: 57600
Data: 8
Parity: none
Stop: 1
Flow control: hardware (or none)
Transmit delay: 0 and 1 communication.
10. In the File menu, select Send File.
11. In the folder
ID programming on the
CD-ROM, select one of the files
7PRG0260 (LEGP).txt and 7PRG0261
(LEHR).txt
.
Select the file corresponding to the type of
collimator currently mounted on the detector.
12. Click Open.
A pop-up appears counting the number of
bytes transferred. When the ID programming
is completed successfully, the pop-up closes
automatically.
13. Close the ttermpro program.
The ECG connection must be restored:
14. Remove the serial interface cable from the
connector A1 RS232 and reconnect the
ECG cable (see step 2 on page 7-12).
15. Remount the detector top cover. See in-
structions in Chapter 6. Repair Proce-
dures.