Philips Joey Dosimeter - Device Management and Support; Battery Status and Recalibration; Product Warranty

Philips Joey Dosimeter
2 pages
Print Icon
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAUTION: We recommend that the display window only be wiped using a soft damp cloth while
exerting very little pressure. Rubbing your ngers across the display window creates static electricity,
which can cause false readings.
Battery
A weak battery will be shown automatically on the digital display – LO BAT.
Recalibration
The Photobilirubin meter is calibrated against lamp standards traceable to the National
Institute of Standards Technology. It is calibrated to be +/-2% or +/- one digit (the least
signicantdigit).
Proper calibration can only be accomplished with a National Institute of Standards traceable
lamp standard. It is recommended that the instrument be returned to Philips Children’s
Medical Ventures for recalibration immediately if the meter has been physically abused (e.g.
dropped, jarred). Otherwise, the recommended time period for recalibration is ANNUALLY.
For recalibration of the Joey Dosimeter measuring device, contact Philips Children’s
Medical Ventures at 1-800-345-6443 or 724-387-4000.
Warranty
Children’s Medical Ventures, LLC. warrants that the Joey Dosimeter will, at the time of shipment,
be free and clear of all liens and encumbrances and will be free from defects in material and
workmanship for a period of one year from date of purchase from Children’s Medical Ventures,
LLC. Children’s Medical Ventures, LLC excludes all other warranties, whether express, implied or
statutory,includinganyimpliedwarrantiesofmerchantabilityortnessforaparticularpurpose.
In no event will Children’s Medical Ventures, LLC. be liable for any incidental or consequential
damages due to breach of this warranty or any other obligation.
1081026
HM 11/08/2010
Manufactured for:
Children’s Medical Ventures, LLC
191 Wyngate Drive
Monroeville, PA 15146 USA
Respironics Deutschland
Gewerbestrasse 17
82211 Herrsching, Germany
www.philips.com/childmed