Joey Dosimeter
Instructions for Use
Description
The Joey Dosimeter is a wide-band spectroradiometer used in neonatal phototherapy.
The response curve of the sensor is between 425nm – 475nm with the peak at 450nm.
The meter measures irradiance in units of microwatts per centimeter per square
nanometer (µw/cm²/nm).
The Joey Dosimeter should only be used to measure the irradiance of the wavelengths
of blue light, 425nm – 475nm, and for no other purpose.
Use with Fiber Optic Phototherapy Devices
1. Placethephotosensorfacedownontheberopticpanel.
2. Press the “PUSH TO READ” button.
3. The Joey Dosimeter measuring device has one scale ranging from 00.0 to 99.0 µw/cm²/nm
(425-475). The light intensity (µw/cm²/nm) can be read directly from the digital display.
NOTE: A “LO BAT” signal will be displayed if the battery needs to be changed.
Output measurements should be taken whenever there is a condition that may affect the
phototherapy light output such as but not limited to:
1. Before each patient use
2. Operating hours of phototherapy device
3. Cleanliness of device’s optical system
4. Condition of light emitting pad.
Battery Information
The Joey Dosimeter operates with a single nine volt (9V) nonrechargeable battery.
Replacing the Battery
1. Remove the battery compartment cover on the lower back side of the Joey Dosimeter.
2. Lift out the 9-volt battery and replace it with a battery of the same type.
3. Replace the battery compartment cover.
Maintenance
Cleaning
Avoidgettingngerprints,grease,ordustonthesensor.Adirtysensorwillresultinfalselow
intensity readings. Clean the sensor with a water dampened lens paper and then wipe dry.
DO NOT immerse in water or apply any water or commercial cleaning agents directly
to the sensor and/or digital display, as this will cause damage to the device.
Question and Answer IconNeed help?

Do you have a question about the Philips Joey Dosimeter and is the answer not in the manual?

Philips Joey Dosimeter Specifications

General IconGeneral
BrandPhilips
ModelJoey Dosimeter
CategoryMeasuring Instruments
LanguageEnglish

Summary

Joey Dosimeter Overview

Device Description

Explains the Joey Dosimeter as a wide-band spectroradiometer for neonatal phototherapy.

Using the Joey Dosimeter

Fiber Optic Phototherapy Device Use

Instructions for connecting and reading measurements with fiber optic phototherapy devices.

Device Operation and Care

Battery Information and Replacement

Details on the 9V battery, low battery indication, and replacement procedure.

Sensor Cleaning and Maintenance

Guidance on cleaning the sensor to ensure accuracy and avoid device damage.

Device Management and Support

Battery Status and Recalibration

Covers low battery display and recommended annual recalibration for accuracy.

Product Warranty

Details the one-year warranty for defects in material and workmanship.