EasyManua.ls Logo

Phymetrix POA200 - Page 21

Default Icon
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
1.4 Calibration
First of all, connect the calibration gas and purge the lines, on the
measuring screen, press the "up/calibration "button, as shown by the
button: ,display the calibration interface,and then adjust the
reading withe UP and DOWN buttons until the analyzer agrees with
the calibration gas value.
Detailed calibration steps:
1. Make sure the Sample/Bypass valve is in the "OFF"position.
2.Allow low oxygen gas flow to purge pressure reducing valve.
3.Allow approximately 1-2min.time for the gas (N
2
would be better)to
purge the lines.
4. Connect the calibration gas into the analyzer sample inlet and set
the flow to about 1 SCFH.
5. Rotate the Sample/Bypass valve to "ON" so gas flows by the
sensor.
6. Wait for a few minutes for the oxygen reading to stablize.
7. Adjust the reading with the UP and DOWN buttons until the
analyzer agrees with the calibration gas value.