Specifications
A-2 LIFEPAK 1000 Defibrillator Operating Instructions
Waveshape and Measured Parameters:
Patient
Impedance ()
I1 (A) I2 (A) I3 (A) I4 (A) T1 (ms) T2 (ms)
Delivered
Energy
(Joules)
25
50.3 20.1 19.7 10.7 5.9 3.9 199
50
28.2 14.6 14.5 9.3 7.5 5.0 204
75
19.8 11.7 11.7 8.2 8.7 5.8 201
100
15.5 10.0 9.9 7.3 9.7 6.5 200
125
12.9 8.7 8.7 6.6 10.4 7.0 201
150
11.1 7.8 7.7 6.2 11.1 7.4 200
175
9.8 7.1 7.1 5.7 11.7 7.8 199
200
8.7 6.5 6.5 5.3 12.2 8.1 199
Note: Table values are nominal for a 200-joule shock.
T1
T2
I1
I2
I3
I4