Specifications
A-4 LIFEPAK 1000 Defibrillator Operating Instructions
Waveshape and Measured Parameters:
Patient
Impedance ()
I1 (A) I2 (A) I3 (A) I4 (A) T1 (ms) T2 (ms)
25 19.4 10.2 10.1 6.6 7.6 5.1
50 13.2 7.4 7.3 5.0 8.1 5.4
75 10.1 5.8 5.7 4.0 8.3 5.6
100 8.3 4.8 4.8 3.3 8.6 5.7
125 7.0 4.2 4.1 2.9 8.8 5.9
150 6.2 3.7 3.7 2.6 8.8 5.9
175 5.5 3.3 3.3 2.3 8.9 6.0
Note: Table values are nominal for a 50-joule shock (200 ÷ 4).