A-6 LIFEPAK 1000 Defibrillator Operating Instructions
ECG Mode
ECG Display Provides nondiagnostic ECG display of the patient’s heart rhythm.
Display
Size (Active viewing
area)
120 mm (4.7 in.) x 89 mm (3.5 in.)
Display Type 320 dot x 240 dot LCD with backlight
Frequency Response 0.55 Hz to 21 Hz (-3 dB), nominal
Waveform Sweep
Speed
25 mm/sec for ECG, nominal
Waveform viewing
time
Minimum 4 seconds
Waveform Amplitude 1 cm/mV, nominal
Display Range Differential: ±1.4 mV full scale, nominal
Heart Rate 20 to 300 BPM digital display.
Display "---" if heart rate is less than 20 BPM.
Heart symbol flashes for each QRS detection
Accuracy: ±4% or ±3 bpm, whichever is greater
Displayed ECG ECG information is received from therapy pads in anterior-lateral
or anterior-posterior positions, or from the 3-wire ECG cable in
Lead II.
Controls
On/Off Controls device power
Shock Controls the delivery of defibrillation energy
Soft Keys Used during device setup and during patient use: Analyze,
Charge, Disarm
Menu button Used to access additional device features
Readiness Display
The readiness display shows device status
OK Indicator Indicates OK when the last self-test was completed successfully.
Battery Capacity
Indicator
Segmented display showing battery capacity
Service Indicator Service required when displayed
Environmental
Note: All performance specifications defined assume that the device has been stored
(two hours minimum) at the operating temperature prior to operation.
Operating
Temperature
0° to 50°C (32° to 122°F)