Electromagnetic Compatibility Guidance
LIFEPAK 1000 Defibrillator Operating Instructions G-3
©2018 Physio-Control, Inc.
The LIFEPAK 1000 defibrillator is intended for use in the electromagnetic environment specified
below. The customer or the user of the LIFEPAK 1000 defibrillator should ensure that the device
is used in such an electromagnetic environment.
Ta b l e 3 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Immunity Test IEC 60601 Test Level Compliance Level
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz outside ISM
bands
1
1. The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur radio bands
between 150 kHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz to 5.4 MHz, 7.0 MHz to
7.3 MHz, 10.1 MHz to 10.15 MHz, 14.0 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to 21.4 MHz,
24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz, and 50.0 MHz to 54.0 MHz.
3 Vrms
6 Vrms
150 kHz to 80 MHz
in ISM and amateur bands
1
10 Vrms ISM
6 Vrms amateur
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m