Electromagnetic Compatibility Guidance
G-4 LIFEPAK 1000 Defibrillator Operating Instructions
SEPARATION DISTANCES
The LIFEPAK 1000 defibrillator is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the LIFEPAK 1000
defibrillator can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
LIFEPAK 1000 defibrillator as recommended below, according to the maximum output power of
the communications equipment.
The LIFEPAK 1000 defibrillator was tested to various RF wireless communication environments
to meet a minimum separation distance as recommended by 60601-1-2:2014 The following table
describes the environments and the equation for calculating the recommended separation
distance at the maximum power level for each band.
Ta b l e 4 Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the
LIFEPAK 1000 Defibrillator
Band (MHz) Service
Maximum
Power (W)
Distance (m) Equation
380-410
TETRA 400
T-GSM-380
T-GSM-410
1.8 0.3
d = 6/27
430-470
GMRS 460
FRS 460
LTE Band 31
4G/LTE-A
2.0 0.3
d = 6/28
470-500 GSM-480 1.8 0.3
d = 6/27
704-787
LTE Band 12, 13, 17, 20, 26
GSM-710
GSM-750
0.2 0.3
d = 6/9
800-960
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5, 8, 27
PRS-900
T-GSM-810
GSM-850
P-GSM-900
E-GSM-900/UTRA
R-GSM-900
T-GSM-900
UTRA Band 5
2.0 0.3
d = 6/28
1480-1530 UTRA Band 11/LPDC (Japan) 2.0 0.3
d = 6/28